59
3.2.1 Trigger Mode
There are three trigger modes in DL-1060: auto, immediate, and single triggering.
You can specify the trigger mode for your measurement. The factory default is auto
triggering when the meter is power-on.
A. Auto Triggering Mode (Front Panel Operation only)
Definition
Auto triggering takes continuous readings at a fastest rate possible for the present
measurement. The rate of taking readings depends on the current settings. This
function is only available through the front panel. The auto triggering is also the
default for trigger mode in DL-1060.
How to use Auto Trigger
1.
Press SHIFT and TRIGGER on the front panel to enable auto trigger mode as
shown in
Figure 3-18
.
※
Note: This operation is only used to return from Single Triggering Mode.
Figure 3-18
B. Internal Immediate Trigger Mode (Remote Interface Operation Only)
Definition
The immediate triggering mode is only available through the remote interface
operations and it is the default for the remote interface operations. When the meter is
in immediate trigger mode, a trigger will be issued immediately as soon as the meter
is in “wait for event” state.