40
To move both cursors at once,
press
X1X2
and then use the
Variable knob.
X1X2
10.00uS
100.0kHz
0.000uV
∆ :
f :
Remove cursors Press Cursor to remove the
onscreen cursors.
Cursor
3-3-2. Using the vertical cursors
Procedure
1. Press the Cursor key.
Cursor
2. Press
X
↔Y
to select the
vertical (Y1&Y2) cursor.
X
↔Y
3. Press
Source
repeatedly to
select the source channel.
Source
CH1
Range
CH1, 2, MATH
4. The cursor measurement results will appear in
the menu.
Parameters
Y1
Voltage level of the upper cursor
Y2
Voltage level of the lower cursor
Y1Y2
The difference between the upper and
lower cursor
V/A
The voltage/current difference (Y1-Y2).
Moving the
vertical cursors
To move the upper cursor,
press
Y1
and then use the
Variable knob.
Y1
123.4mV
To move the lower cursor,
press
Y2
and then use the
Variable knob.
Y2
12.9mV
To move both cursors at once,
press
Y1Y2
and then use the
Variable knob.
Y1Y2
1 0 . 5 m V
Remove cursors Press Cursor to remove the
onscreen cursors.
Cursor