164
5. Press
Vertical
on the side menu.
6.
A message appears to “Now performing vertical
calibration…
CH1
Connect the CAL output to channel, then press
the Vertical key”.
7. Connect the calibration signal from the rear
panel to the Channel 1 input with a BNC cable.
It will use very low level signal, please use the
cable with the high shield effect and short.
CAL
CH1
8. Press
Vertical
again after
connecting CAL to the channel 1
input.
The calibration for Channel 1 starts and ends
automatically, in less than 5 minutes.
A message is displayed when the calibration
procedure has ended.
9. Repeat the above step for Channel 2, 3* and 4*
when prompted.
*4 channel models only.
10. When the calibration for all channels has
completed, the display goes back to the default
state.
9-1-3. Probe Compensation
Panel Operation 1. Connect the probe between the Channel 1 input
and the probe compensation output (default set
as 2Vp-p, 1kHz square wave) on the front
panel. Set the probe attenuation to x10.
2. Alternatively, the probe compensation
frequency can be changed. See page 117 for
details.