Table 1. ADS1x98ECG-FE Default Jumper/Switch Configuration
Jumper
Default Position
Description
JP1
Installed
RLD feedback
JP2
Installed 1-2
AVDD selected for bipolar supply operation selected (AVDD = +2.5V)
JP3
Header Not Installed
External Vref buffer not connected
JP4
Installed
EVM +5V provided from J4 (power header)
JP5
Open
PWDN pin controlled from J5 header (pulled up to DVDD)
JP6 to JP14
Header Not Installed (Pins 1-2
shorted on PCB)
DC-coupled input signals
JP15
Installed 2-3
Shield drive is open
JP16
Installed
Wilson Central Terminal (WCT) connected to INM for CH1 and CH4-8
JP17
Header Not Installed
ECG shield drive
JP18
Installed 2-3
CLK
connected to OSC1
JP19
Installed 1-2
OSC1 enabled
JP20
Installed 2-3
AVSS selected for bipolar supply operation (AVSS = -2.5V)
JP21
Installed 1-2
CS
connected to DSP via J3.1
JP22
Installed 2-3
START
comes from J3.14
JP23
Installed 1-2
CLKSEL
set to 0 (ADS1198 uses Ext Master Clock (OSC1))
JP24
Installed 2-3
DVDD supply = 3.3V
JP25
Header Not Installed
No external reference selected
JP26
Installed 1-2 (top)
WCT connected to CH8- input
Installed 3-4 (bottom)
ECG_V1 connected to CH8+ input
JP27
Installed 1-2 (top)
WCT connected to CH7- input
Installed 3-4 (bottom)
ECG_V5 connected to CH7+ input
JP28
Installed 1-2 (top)
WCT connected to CH6- input
Installed 3-4 (bottom)
ECG_V3 connected to CH5+ input
JP29
Installed 1-2 (top)
WCT connected to CH5- input
Installed 3-4 (bottom)
ECG_V4 connected to CH6+ input
JP30
Installed 1-2 (top)
WCT connected to CH4- input
Installed 3-4 (bottom)
ECG_V2 connected to CH4+ input
JP31
Installed 1-2 (top)
ECG_RA connected to CH3- input
Installed 3-4 (bottom)
ECG_LL connected to CH3+ input
JP32
Installed 1-2 (top)
ECG_RA connected to CH2- input
Installed 3-4 (bottom)
ECG_ LA connected to CH2+ input
JP33
Installed 1-2 (top)
WCT connected to CH1- input
Installed 3-4 (bottom)
ECG_V6 connected to CH1+ input
2.2
ADS1x98ECG-FE Operation
To prepare to evaluate the ADS1298 with the ADS1298ECG-FE, complete the following steps:
1. Verify the jumpers on the ADS1298ECG-FE are as shown in
(note that these settings are the
factory-configured settings for the board).
2. Verify that the jumpers on the MMB0 motherboard are configured as shown below:
•
MMB0 J13A
→
Open
•
MMB0 J13B
→
Open
•
Refer to
for details about the MMB0 power supply.
3. Install the ADS1298ECG-FE software using the latest software version. The latest software can be
downloaded from the
or
. Double click
the installer and follow the instruction to complete the software installation. For detailed installation
information and screenshots, see