6 Using the product
65
Reset hour meter (NO sensor only)
1.
[Reset].
2.
Confirm confirmation request:
Yes
→
[OK]
6.1.4.6.
Calibration data
Use this function to display the current calibration data (comparison
of nominal value and actual value without sensor adjustment).
Calling up the function:
>
[ ]
→
Sensor settings
→
[OK]
→
Calibration data
→
[OK]
.
Options
>
[Options]
→
[Print]
: The current calibration data of all sensors
are printed out.
6.1.4.7.
Adjustment data
Use this function to display the current adjustment data and the
sensor status of the individual sensors.
The condition of the sensor is checked every time the sensor is
adjusted. The graph view shows the last 25 adjustments.
Call up function:
>
[ ]
→
Sensor settings
→
[OK]
→
Adjustment data
→
[OK]
Options
>
[Options]
→
[Print]
: the current adjustment data for all sensors
is printed out.
>
[Options]
→
[Graphic]
: the status of the selected sensor is
graphically displayed.
Threshold Explanation
100%
Full capacity
70%
Reduced sensor sensitivity.
Recommendation: Acquire a replacement
sensor
50%
Replace sensor
6.1.4.8.
Negative value
The display for negative values can be activated / deactivated.
Calling up the function:
>
[ ]
→
Sensor settings
→
[OK]
→
Negative values