Automatic
Vicat Needle Apparatus
1.0306
13
If button is pressed while the sys-
tem is performing the normal test:
This will stop the normal test procedure.
3.3 The test procedure
Insert the drop rod from above into the device (in the reverse direction as shown in
the drawing below).
Switch on the system. The drop rod will now move into the upper position.
Make your selection for the following parameters:
o
The standard that will apply to the test
o
The type of fall (free fall or damped fall)
o
The desired time
Place the prepared test specimen on the rotary disk.
Press the START button. The test procedure will then begin.
Programm
Needle length
[mm]
Norm
Vicat-Ring
[Size min.]
Anzahl der
Einstiche
A
50
EN 196
Ø 70/80
26
B
50
ASTM 191
Ø 65/75
37
C
50
EN 196
Ø 70/80
90
D
30
Erstarrungs-
ende
Ø 70/80
22
Test + 5’
50
Normsteife
1
G
50
(Konus)
Gips
DIN 1168
Ø70/80
16