background image

 

6/8 

4

 

Product Specification 

 

Product 

TSR4 WLAN Wireless and USB Radon Probe 

Type symbol 

042 127 210 000 

Average measurement sensitivity 

0,25 count/hour/Bq.m-3 
(method RaA+RaC; 

15°C ÷ 30°C; rel. hum. 20% ÷ 40%)

 

Measuring range 

MDA 

 100 000 Bq/m

3

;  

MDA = 100 Bq/m

3

 per 1 hour or 20 Bq/m

3

 per 24 hours 

Measurement uncertainty 

< 13% at 300 Bq/m

3

 per 1 hour; 

< 3% at 300 Bq/m

3

 per 24 hours 

Measuring chamber capacity 

0,176 dm3 

Response rate 

< 30 minutes (RaA); < 3 hours (RaA + RaC) 

Radon records 

calculated from RaA (quicker, less sensitive) 

 

calculated form RaA + RaC (slower, more sensitive) 

Measuring relative humidity range 

10 

 90 % 

Measuring temperature range 

-

20 to + 60 °C 

 

Radio interface 

868MHz  

Max number of measuring network 
elements 

16  

Probe to terminal unit distance (RF range)  depends on number of walls and building material, up to 600 m 

in open space 

Possibility of using repeater for RF range 
extending 

yes 

Results reading interval (from probe) 

240-65535 sec (4 min - 18.2 hours) 

Records saving interval (probe) 

1- 255 minutes, default 1 hour 

Results memory capacity in probe 

4096 (150 days of 1 hours records) 

Powering 

internal rechargeble accumulator; charging via USB 

Accu life after full charging 

>1 year 

Current radon concentration results  

short-term (0,5 hour running average from RaA) 

 

long-term (24 hours running average from RaA + RaC)

 

Dimension 

Ø 80 x 

175 mm 

 
 

5

 

Operating Instructions 

 
Switching on and off: 

   The probe measures radon concentration autonomously and communicates in wireless network only if 

the switch is in position 

„I“( 

switch on). 

The switching on is signalized by LED diode „STAT“ according  chart 

below.  

 If the switch is in position 

„0“ (

switch off) the probe doesn

´

t measure radon concentration and doesn

´

communicate in wireless network.  In switching off mode the probe only keeps running real time for correct 
date  and  time  of  records  in  case  of  switch  it  on  again.  By  switching  off  the  probe  doesn

´

t  lose  previous 

records of measurement. 

The switching on is signalized by LED diode „STAT“ according  chart below. 

  

It is possible to download data from probe over USB in both position of switch.   

 

Summary of Contents for TSR4

Page 1: ...cal Specifications Operation Manual v 1 2017 Table of Contents 1 Introduction 2 2 Description and Utilization 2 3 Scope of Delivery 5 4 Product Specification 6 5 Operating Instructions 6 6 Basic Maint...

Page 2: ...d product If you experience any problems with your product please contact us at TESLA Rubeska 215 1 190 00 Prague 9 Vysocany www tesla cz 1 Introduction This document describes technical specification...

Page 3: ...battery power portable radon measuring probe supports flexible placing options within monitored structures Accumulator will last for more than 1 year after full charging After switching on TSR4 immedi...

Page 4: ...on and on the basis of measured set concentration level value it sends command to actuator wireless actuator or actuator in central unit can be use which is hardwired with power relay Power relay swit...

Page 5: ...ords Quicker download data from probes if you download data after end of the measurement Possibility of various identification writing into internal memory of probe Compatibility TSR4 and TSR3 radon p...

Page 6: ...n 18 2 hours Records saving interval probe 1 255 minutes default 1 hour Results memory capacity in probe 4096 150 days of 1 hours records Powering internal rechargeble accumulator charging via USB Acc...

Page 7: ...way of radon probe If the probe is used for autonomous measurement and data downloading via USB interface then the setting and configuration are realized by usbRADONview application UsbRADONview appl...

Page 8: ...n of Conformity will be delivered by Tesla producer on request If interested please use contacts on the web www tesla cz 8 Repairs Any repairs and non basic maintenance must be performed exclusively b...