Product Specification
Product
TSR4 WLAN Wireless and USB Radon Probe
Type symbol
042 127 210 000
Average measurement sensitivity
0,25 count/hour/Bq.m-3
(method RaA+RaC;
15°C ÷ 30°C; rel. hum. 20% ÷ 40%)
Measuring range
MDA
–
100 000 Bq/m
3
;
MDA = 100 Bq/m
3
per 1 hour or 20 Bq/m
3
per 24 hours
Measurement uncertainty
< 13% at 300 Bq/m
3
per 1 hour;
< 3% at 300 Bq/m
3
per 24 hours
Measuring chamber capacity
0,176 dm3
Response rate
< 30 minutes (RaA); < 3 hours (RaA + RaC)
Radon records
calculated from RaA (quicker, less sensitive)
calculated form RaA + RaC (slower, more sensitive)
Measuring relative humidity range
10
–
90 %
Measuring temperature range
-
20 to + 60 °C
Radio interface
868MHz
Max number of measuring network
elements
16
Probe to terminal unit distance (RF range) depends on number of walls and building material, up to 600 m
in open space
Possibility of using repeater for RF range
extending
yes
Results reading interval (from probe)
240-65535 sec (4 min - 18.2 hours)
Records saving interval (probe)
1- 255 minutes, default 1 hour
Results memory capacity in probe
4096 (150 days of 1 hours records)
Powering
internal rechargeble accumulator; charging via USB
Accu life after full charging
>1 year
Current radon concentration results
short-term (0,5 hour running average from RaA)
long-term (24 hours running average from RaA + RaC)
Dimension
Ø 80 x
175 mm
5
Operating Instructions
Switching on and off:
The probe measures radon concentration autonomously and communicates in wireless network only if
the switch is in position
„I“(
switch on).
The switching on is signalized by LED diode „STAT“ according chart
below.
If the switch is in position
„0“ (
switch off) the probe doesn
´
t measure radon concentration and doesn
´
t
communicate in wireless network. In switching off mode the probe only keeps running real time for correct
date and time of records in case of switch it on again. By switching off the probe doesn
´
t lose previous
records of measurement.
The switching on is signalized by LED diode „STAT“ according chart below.
It is possible to download data from probe over USB in both position of switch.