5/6
Power supply:
According to operation method of device can be supplied:
1)
By internal accumulator for portable use
–
Device includes internal accumulator which is able to
ensure autonomous operation of device for more than 12 months without charging. Depends on
frequency of wireless data retransmiting and depends on climatic condition of device use.
Accumulator is charged over USB port and over provided USB cable. The USB cable is possible to
connect to PC or to delivered power adapter. Status of accumulator and charging process are
described in paragraph ´Basic Maintenance/Accumulator charging´
2)
By mains power supply 230V/50Hz for stationary use
–
Device is permanently supplied by delivered
power adapter 230VAC/5VDC. Power adapter is connected to device via provided USB cable. In
case of blackout the internal accumulator ensures UPS function.
Configuration:
Setting and configuration of Repeater and whole measuring system is realized by Central Unit connected
to PC and TERAview application.
Central Unit is not included with package of TRR3 and it is sold and
delivered extra; see
TERAview application, drivers and user manual with detail configuration description is free downloaded
For successful repeater configuration in measuring system it is necessary to know repeater radio
channel number (communication wireless channel) and P2P address (identification in wireless net). Both
parameters are printed out on repeater serial number plate. Repeater radio channel number is possible to
change by TERAview application and it must be identical to central unit radio channel number. P2P address
is permanent and it occur in one big wireless net only once. Only P2P address of Central Unit can be
identical to P2P address of other elements in network.
6
Basic Maintenance
Accumulator charging:
During portable use of device is essential to monitor state of internal accumulator and recharge it if
necessary. If accumulator is discharged the device automatically turns off.
Current state of accumulator can be determined by these ways:
1)
By LED diode ´STAT´
- If LED starts blinking in yellow color it indicates low voltage of accumulator.
(see paragraph “Operation Manual / LED diode “STAT””)
2) In TERAview application - where you can check current accumulator voltage. Voltage should not
fall below 3.5 V, in limit conditions falls below 3.3V.
Accumulator is charged via USB port using supplied USB cable. USB cable can be connected to PC or to
supplied power adapter. Connect USB cable with power to USB port of device
. LED diode ´CHRG´ n
ext to
USB port of device indicates charging status according to following chart:
LED diode ´CHRG´
Color
Description
Green
Accumulator is fully charged
Yellow
Accumulator is being charged
Green
-
Yellow
alternate blinking
Accumulator is damaged, contact Service Center
No light, No blinking
It is not connected to an external power supply or device is damaged.
Accumulator is fully charged when LED diode ´CHRG´ is green. You can disconnect USB cable.
7
Repairs
Any repairs and non basic maintenance must be performed exclusively by Tesla manufacturer.
TESLA
Podebradska 56/186
180 66 Prague 9