39
Use F5 to return to the original menu level.
5.12 Storing voltage settings
Pre-programmed discharge voltage values can be stored in four memory loca-
tions for both air and contact-discharges. As delivered, the instrument has the
test levels set according to IEC/EN 61000-4-2, Ed. 1.2:2001.
The values stored in memory can be altered arbitrarily.
Select discharge mode (air-discharge or contact-discharge) with F1.
Call up «Voltage» submenu with F2.
Level
Test voltage
contact-discharge
Test voltage
air-discharge
1
2
3
4
2 kV
4 kV
6 kV
8 kV
2 kV
4 kV
8 kV
15 kV