5.6 RS-232C Commands in Detail
5.6.1 System Queries
MODEL?
COUNT?
MODEL?
A query to see the model name, firmware version, and the date of manufacture.
Type
Query
Input
MODEL?
Response
<Name>_<Version> <Date>
Example
MODEL?
TC-5972DP_STD_Ver: 5.2 2019.04.09
COUNT?
A query to check how many times the lid has opened.
Type
Query
Input
COUNT?
Response
Numeric 8 digits (ex, 00000029)
5.6.2 Configuration Commands
5.6.2.1 Handler & Probe Configuration Commands
Shield Box will reboot whenever the lid or fixture operation mode is changed, resulting in
opening the lid. If the lid is already open, it will close and then open amid the motion.