136
Non-Signaling Test
MTP200B does not support USB HCI.
BT LE Direct Test Mode (DTM) test can be performed by remote control program provided separ
ately.
21.2 Non-Signaling Test
If DUT can be controlled through DUT control software provided from Chipset company, not using HCI
interface and 2-Wire interface, it is to be called Non-Signaling Test.
DUT’s TX Test (Test for DUT’s transmitting efficiency)
By using the Test Control Software provided by the Chipset company, set DUT to transmit the LE
packet required for test and measure DUT’s TX efficiency with MTP200B.
Figure 21-2
Configuration for BLE TX Measurement Using a Shield Box