122
Signal Generator
2.
MTP200B Setting
A. FCN -> Recall -> RESET <Resets the device>
B. MODE: GEN -> MODE -> CW <Set the measurement mode CW>
C. FREQUENCY: 2400 MHz <Input the CW frequency>
D. POWER: -5 dBm
E. RUN: ON,
3.
Configure the frequency counter as follows:
A. Frequency Counter: Press
Freq Ch 2
key
B. E4440A: [Mode] -> Measuring Receiver -> Frequency Counter key
4.
Record the measurement of the Frequency Counter or E4440A and compare with the specificatio
n.
Table 19-7
Frequency Accuracy
Item
Measurement Result [MHz]
Frequency Counter Measurement Value
19.3.2 CW Mode Power Generator Level Accuracy
Specification
: ±1 dB
Objective
: Measure the power level accuracy of the output signal.
1.
Test Setup: Connect the device as shown below.
2.
MTP200B Setting:
A. FCN -> Recall -> RESET <Resets the device>
B. MODE: GEN -> MODE -> CW <Set the measurement mode CW>
C. FREQUENCY: 2402 MHz <Input the CW frequency>
D. RUN: ON, POWER : -5 dBm
Summary of Contents for MTP200B
Page 1: ...MTP200B WLAN BT LE Tester Operating Manual R20160511 ...
Page 10: ...x ...
Page 11: ...xi ...
Page 12: ......
Page 14: ......
Page 16: ......
Page 18: ......
Page 30: ...14 GEN Generator Menu Tree 4 2 GEN Generator Menu Tree Figure 4 2 GEN Generator Menu Tree ...
Page 31: ...ANL Analyzer Menu Tree 15 4 3 ANL Analyzer Menu Tree Figure 4 3 ANL Analyzer Menu Tree ...
Page 34: ......
Page 41: ...PART 2 WLAN Test 25 ...
Page 42: ......
Page 62: ......
Page 63: ...PART 3 BT LE Test 47 ...
Page 64: ......
Page 68: ......
Page 76: ......
Page 82: ......
Page 83: ...PART 4 Programming Guide 67 ...
Page 84: ......
Page 102: ......
Page 103: ...PART 5 Command List 87 ...
Page 104: ......
Page 129: ...PART 6 User s Device Diagnosis 113 ...
Page 130: ......
Page 144: ......
Page 145: ...PART 7 Appendix 129 ...
Page 146: ......
Page 156: ......