Instruction manual for MH & MH+M
151
You’ll find on the page :
•
Part status
- Good
- Not good
- Program end (if the sequence does not include any tolerances)
•
The number of
-
Measurements in black (measurements whose tolerances have not been entered)
-
Values in tolerances in green
-
Values to be reworked in yellow
-
Values out of tolerances in red
•
Run time
•
Run number
•
Part name (corresponding to program name)
•
Batch name
•
Operator name
It is possible to filter the result values by status. To do so, press one of the
button. It will
display only the corresponding values having the same status (all good, all bad or all having
to be reworked).
22.12 To run a
sequence in
loop
Once the end of a sequence run is reached, the global results page is displayed. You can
either re-run the same sequence from this page using
option or exit the process with
22.13 Calibration
block and
execution in a
loop
A sequence that has been done from ST2 mode will, most of the time, include a probe
calibration block as first block. When running a sequence having such block as first one (if
there’s no other probe calibration block in the same sequence) running the sequence for the
second time will imply that the calibration block won’t be taken into account.
22.14 Reference
block and
execution in a
loop
If a measurement sequence takes only one reference (a single reference block) into account
during its execution, the software will offer you two ways to manage it:
•
Measuring the reference at each run of the sequence
•
Measuring the reference only at the first sequence run. This reference is kept in memory
for future executions (under the condition of not leaving the sequence execution mode).