RS
PrintingHumidity/Temperature Meter
1362
INSTRUCTION MANUAL
Page 1: ...RS Printing Humidity Temperature Meter 1362 INSTRUCTION MANUAL...
Page 2: ...MATION 1 II SPECIFICATIONS 2 III LOCATION OF PARTS 4 IV GENERAL PRECAUTIONS AND PREPARATIONS FOR MEASUREMENT 8 V OPERATING PROCEDURE 8 VI SETTING MODE 10 VII INTERVAL PRINTING SET UP 14 VIII BATTERY R...
Page 3: ...be impaired Environmental conditions c Altitude up to 2000 metres d Indoor use only e Relatively humidity 95 max f Operating temperature 0 to 50 32 to 122 Maintenance clearing c Repairs or servicing...
Page 4: ...rs interval printing 60 minutes excluding beeper 100 hours without printing beeper Printing interval 6 sec to 59 min 59 sec Operating temperature and humidity 0 to 50 32 to 122 below 95 RH Storage tem...
Page 5: ...8 5 47 2 to 137 3 Accuracy 1 0 1 8 Temperature T2 External thermocouple sensor Range 200 to 1333 328 to 2431 Resolution 0 1 from 200 to 999 0 1 from 328 to 999 9 1 or 1 outside this range Accuracy 50...
Page 6: ...points are displayed on the screen 2 Power on off button 3 Button for selecting temperature units or 4 PRINT c Press this button to start printing the current data The data format is as follows LINE...
Page 7: ...the printer paper one line d Force the printer to stop printing and clear any data in the print buffer 7 DEW POINT Press this button change the display from humidity to dew point Fig 2 8 MAX MIN c Pre...
Page 8: ...ours minutes 11 DOWN Press this button to decrease the value of the selected parameter 12 RIGHT Button for selecting the next desired parameter for setting 13 UP Press this button to increase the valu...
Page 9: ...indicator 3 Low battery indicator 4 Printing indicator 5 Interval printing start time setting indicator 6 Interval printing stop time setting indicator 7 Calendar time indicator 8 Interval printing ti...
Page 10: ...er adapter ensure it is suitable has the correct polarity and is undamaged 4 Ensure the battery cover is correctly fitted otherwise measurement errors may occur 5 If the instrument is to be stored for...
Page 11: ...K type thermocouple if connected 4 Printing data c Press the PRINT button once to print one set of data as shown on the LCD d If continuous printing is required press and hold down the PRINT button u...
Page 12: ...continuous interval printing mode and begin printing Press the PRINT or FEED button to exit printing mode The interval printing function works only when the start printing time is earlier than the sto...
Page 13: ...g When in setting mode the adjustable parameter will flicker until setting is finished Press the or button to select the required parameter to set then press the or button to increase or decrease the...
Page 14: ...nting start time minutes value fig 7 5 Setting the printing start time hours value fig 8 6 Setting the printing stop time minutes value fig 9 7 Setting the printing stop time hours value fig 10 8 Sett...
Page 15: ...13 fig 11 9 Setting the current time hours fig 12 10 Setting the calendar date fig 13 11 Setting the calendar month fig 14 12 Setting the last two digits of the calendar year...
Page 16: ...he required seconds and minutes from 00 06 to 59 59 3 Select the required time to start printing Set the minutes and hours from 00 00 to 23 59 4 Select the required time to stop printing Set the minut...
Page 17: ...humidity probe and any K type thermocouple Remove the battery cover 2 Remove batteries from the holder and replace with qty 6 new 1 5V size AAA alkaline batteries 3 Refit the battery cover and check...