6
Cautions
•
Before removing or inserting the catheter through the sheath,
aspirate blood from the 3-way stopcock to remove any fi brin
deposition which may have accumulated in or on the tip of
the sheath (Fig. 8).
•
When puncturing, suturing, or incising the tissue near the
sheath, be careful not to damage the sheath. Do not put a
clamp on the sheath nor bind it with a thread.
The fl ushing line may also be used as a continuous infusion site by
connecting an infusion line to the 3-way stopcock (Fig. 9).
12. When inserting, manipulating or withdrawing a catheter from the
sheath, always hold it in place. To temporarily suture the sheath (for
continued access) use the suture eye.
Cautions
•
Do not place suture on the sheath tubing since this may
restrict access/fl ow through the sheath.
•
Do not use a power injector through the side tube and 3-way
stopcock.
13. When exchanging catheters, remove the used catheter and repeat
Step 11.
14. After the intended procedure is completed, remove the catheter
and then the sheath.
Fig. 8
Fig. 9
3-Way Stopcock
Catheter
Infusion Line
CAUTION FOR STORAGE
Do not store at extreme temperatures and humidity. Avoid direct
sunlight.
PM-01979