Bionova® MiniBio
Page 9
Instructions for use
An audible alarm will sound every time a positive result
is detected in the Bionova® MiniBio incubator. The alarm
allows the user to immediately detect a positive result
without the need to visually control the equipment. The
alarm can be cancelled by pressing the button .
When a BI is removed from its position during incubation
process, an audible alarm will sound to indicate that the
tube should be returned to the corresponding position.
If the indicator is not re-introduced into the reading posi-
tion after 10 seconds of being withdrawn, the reading will
be cancelled and the cancelation will be informed throu-
gh the printing of a ticket.
Cancellation of a reading
Audible alarm
The Bionova® MiniBio incubator contains an internal tem-
perature control. If the temperature falls outside of the spe-
cified range (37±2) ºC or (60±2) ºC, the blue temperature
led light will start to twinkle.
Temperature can be externally monitored by using a Bio-
nova® TB-IC1020 thermometer or other similar. The Au-
to-Reader MiniBio has a special hole for placing the ther-
mometer which is located at the right side of the device.
Temperature monitoring
Clean the outer surface of the Bionova® MiniBio incubator
with a damp cloth moistened with a small amount of de-
tergent. Do not clean interior components. If additional
cleaning of the internal components of the equipment is
required, contact your distributor or manufacturer.
Care and cleaning
Replacement of the paper roll
Thermal paper specifications
Recommended paper: JUJO AF50KSE3 or similar (order
code ICTP).
Paper width : 58 mm
Maximum paper thickness: 60 g/m2
Maximum diameter size: 23 mm
The printer uses thermal paper rolls.
1. To replace the paper
roll, pull the handle of the
printer’s door. Open the
printer’s cover and remo-
ve the spent roll.
2. Place the new paper roll
with the outer side up.
3. Close the printer cover
by pressing on the sides
of the lid.
NOTE:
You can check the correct paper replacement by
pressing
, this button will force the advance of the
paper a few centimeters.
The incubator allows reprinting of the last 3 results by
fluorescence readout. For printing the results, press the
button
for 3 seconds.
Reprinting of results
Summary of Contents for BIONOVA MiniBio
Page 1: ...Rev 5 September 2018 MiniBio Innovation in technologies for Infection Control ...
Page 2: ......
Page 14: ......
Page 15: ...Rev 5 Septiembre 2018 MiniBio Innovación en tecnologías para el Control de Infecciones ...
Page 27: ......
Page 28: ......
Page 29: ...Rev 5 Setembro 2018 MiniBio Inovação em tecnologias para Controle de Infecção ...
Page 41: ......
Page 45: ......
Page 46: ......
Page 47: ......
Page 48: ......