background image

Explosion-Proof

 

RSR

 

Fan/Filter Unit

       

Quick-Start Operating Guide    

Doc . 1788-12

 

© 

Copyright 2015 Terra Universal Inc.   

All rights reserved 

Revised November 2019 

 

 

 

Terra Universal, Inc.  •  TerraUniversal.com  •  800 S. Raymond Ave.  •  Fullerton, CA  92831  •  TEL: (714) 578-6000  •  FAX: (714) 578-6020 

1

 

IMPORTANT SAFETY INSTRUCTIONS 

 

READ AND SAVE THESE INSTRUCTIONS 

Proprietary Notice

 

This manual pertains to proprietary 

devices manufactured by Terra 

Universal, Inc. Neither this 

document nor any portion of it may 

be reproduced in any way without 

prior written permission from Terra 

Universal.

 

Safety Notice

 

A thorough familiarity with all operating guidelines is 

essential to safe operation of the product.  Failure to 

observe safety precautions could result in poor 

performance, damage to the system or other 

property, or serious bodily injury or death. The 

following symbols are intended to call your attention 

to two levels of hazard involved in operation.

 

Terra Universal makes no 

warranties applying to information 

contained in this manual or its 

suitability for any implied or inferred 

purpose.  Terra Universal shall not 

be held liable for any errors this 

manual contains or for any 

damages that result from its use. 

 

CAUTION

 

Cautions are used when failure 

to observe instructions could 

result in significant damage to 

equipment.

 

 

WARNING

 

Warnings are used when failure 

to observe instructions or 

precautions could result in 

injury or death.

 

The information presented here is subject to change without notice.

 

1.0

 

Introduction 

This manual documents installation and operation of Terra Universal’s Room-Side 

Replaceable (RSR) Explosion-Proof WhisperFlow™ Fan/Filter Unit (FFU). The FFU 

provides uniform, laminar flow, HEPA-filtered air to the area below the mounted unit.  

2.0

 

Operation 

The unit consists of a ¼-HP, explosion-

proof motor driving a forward-curved 

centrifugal blower.  Air is drawn into the 

unit through a pleated pre-filter and in 

turn to an insulated chamber where it is 

then directed to an exit HEPA filter with 

a built-in diffuser.  
The unit can also be equipped with a 

3/8”-diameter challenge port, shown in 

photo to the right. Take differential 

pressure measurements to monitor filter performance, and insert aerosols for leak-test 

certification. Port is capped when not in use. 

3.0

 

Installation 

The RSR WhisperFlow FFU is designed to allow filter replacement from the inside of the 

cleanroom, without breaching the seal. A gel-sealed filter can be separated from the fan 

module for replacement. The fan module stays in place on top of a standard 2’x4’ clean 

room ceiling grid and includes a threaded fixture at each corner to accommodate four 

eye-bolts (not provided) that can be used to assist in positioning the unit. 
All metal components are grounded. A grounding wire is clipped to metal hardware on the 

filter screen as well as the metal threads running through the pre-filter mesh. Make certain 

that these wires are in place before final installation. 

WARNING:

 

To reduce the risk of fire, electric shock, or injury to persons, do not 

use this fan with any solid-state speed control device, and observe the 

following:

 

A)

 

Use this unit only in the manner intended by the manufacturer. If you have questions, 

contact the manufacturer. 

B)

 

Before servicing or cleaning unit, or replacing a filter, switch power OFF at the 

service panel and lock the service disconnecting means to prevent power from being 

accidentally switched on. When the service disconnecting means cannot be locked, 

securely fasten a prominent warning device, such as a tag, to the service panel.  

C)

 

When the removal/disconnection of either filter is required due to service or 

component replacement, they are to be remounted as previously installed. 

D)

 

Installation work and electrical wiring must be done by qualified person(s) in 

accordance with all applicable codes and standards, including fire-rated construction. 

All metal components must be properly grounded. 

E)

 

When cutting or drilling into a wall or ceiling, do not damage electrical wiring and 

other hidden utilities.  

4.0

 

Start up  

WARNING: 

Keep unit disconnected from power supply during inspection 

Prior to providing power to the unit, check that no damage has occurred during shipping. 

This can be accomplished via a visual check to make sure there are no visible dents or 

penetrations. If the unit is intact, unclip the grounding wire and remove the pre-filter. 

Manually rotate the fan wheel to make sure it is not in contact with any stationary parts 

and that there are no visible loose screws or bolts obstructing the wheel’s rotation. Once 

visual inspection is completed, have a professional perform the installation who is familiar 

with Class I, Groups C & D and Class II, Groups F & G hazardous location requirements. 

NOTE: Remember to replace the filter-screen and pre-filter grounding wires. After 

installation, with power applied, the fan will rotate, and filtered air will exit the HEPA filter. 

5.0

 

Cleaning and Maintenance 

WARNING:

 

Disconnect from power supply before servicing unit or replacing 

filters. When servicing or replacing either filter, the new filter is to be 

installed in the same manner as the filter it replaces.

 

The scheduled maintenance of the unit depends on the installed location and consists of 

cleaning or changing the pre-filter and the HEPA filter. It is recommended that the pre-

filter be inspected and cleaned every three months or sooner depending on the 

cleanliness of the external environment. The HEPA filter cannot be cleaned and must be 

replaced when the laminar flow rate falls below 70 feet/min.   

 

Explosion-proof FFU to be installed with wiring embedded in filled rigid conduit. 

 

5.1 Replacing the RSR Filters

 

The RSR filter is designed so that the HEPA or ULPA filter can be replaced from inside 

the cleanroom without breaching the controlled environment. The filter assembly is 

detached from the fan module, which stays in place in the ceiling grid. Follow these 

instructions to replace the filter. 
First, you must remove the outer screen covering the filter that is held in place with friction 

catches (Photo 1). This screen protects the filter from damage and enhances airflow 

uniformity. While gently pulling on the screen to disengage the catches, detach the 

grounding wire clip as shown in Photo 2. The clip is attached to a steel rivet holding one 

of the catches in place. 

 

Photo 1

: Friction catches hold the screen to the fan module housing. 

 

Pre-Filter 

Lid 

Assembly 

HEPA/ULPA 

Filter 

Grounding 

Wire 

Optional 

Challenge 

Port 