1. IR Sensor
2. LCD Display
3. Mode selection: body/ambient
4. Down Button
5. Up Button
6. MODE Button
7. Measurement Trigger
8. Battery Cover
7. DEVICE CONSTRUCTION
4
DT-8806
PYROMETER
Version 2.0
USER MANUAL / WARRANTY
8. CONFIGURATION
I. Measuring mode
The Non-contact Body Infrared Thermometer is specially designed to take the body temperature of
a human being. For this, use the Body mode. Measurement range for Body mode: 32 - 42,5 C (86 -
108 F).
You can also use the thermometer to measure the temperature of an area or an object, food, liquid,
or room temperature. For this, use the Surface mode. Measurement range for Surface mode: 0 - 60
C (32 - 140 F).
II. Choosing The Temperature Unit °C / °F - F1 function
Press “
MODE
” button for 2 seconds, the screen displays: F1. Select “
DOWN
” for Celsius degrees
or “
UP
” for Fahrenheit degrees.
III. Alarm Setup - F2 Menu
Press “
MODE
” touch for 2 seconds, the screen displays: F1. Press twice “
MODE
” touch to get F2.
Select “
UP
” to increase the threshold by 0.1 °C (0.1 °F) or “
DOWN
” to deuce it by 0.1 °C (0.1 °F).
The alarm threshold default is 38 °C (100.4 °F).
IV. Pyrometer display calibration/Total Difference (This function is only effectictive in Body
mode - body temperature measurement)
To adjust the total variation of your Non-Contact Body Infrared Thermometer, press “
MODE
” button
for 2 seconds, the screen displays: F1. Press “
MODE
” button two times to get F3. Select “
UP
” to
increase the difference by 0.1 °C (0.1 °F), “
DOWN
”
to reduce it by 0.1
°C (0.1 °F). Temperature
value can be changed in: -3,0...3,0 °C range. In the cases of seasonal or environment changes a
verification and adjustment should be carried out.
V. Buzzer ON/OFF - F4(F3) Menu
Press
“
MODE
”
button for 2 seconds, the screen displays: F1. Press three(two) times
“
MODE
”
Button to get F4(F3).
Select “
UP
” to open the buzzer ( asound icon is displayed on the LCDC screen), press “
DOWN
” to
stop it ( the icon dissapear).
°
°
°
°