VT-R-01, VT-R-03 heating constant temperature baths 

Operating manual 

 

USING THE HEATING CONSTANT TEMPERATURE BATHS  

 

Before using bath circulator, carefully read the operating manual. 

Preparation 

 

Carefully  select  a  spot  for  installing  in-

strument  with  free  air  access  for  the  circulator 

ventilation.  Make  sure  it  is  far  away  from  any 

kind of heat source. 

Place the instrument on an even surface 

with a pad, made of nonflammable material. 

Insert output pipe 2 with its taper end up 

to the stop in corresponding socket 1 and secure 

it with a screw. 

Install  circulator 3  into  the  adapter  on 

the  bath  cover 5.  Secure  the  circulator  with 

screws 4. 

 

While  setting  up  the  circulator,  supply  cord 

should not be connected to the power source. 

Attach a pressure gauge to a Reid-vapor 

pressure vessel. Insert the vessel into adapter 6 

and slide into the slot. 

 

 

Fill  up  the  bath  tank  with  thermal  fluid 

through  aperture  under  bath  cover 5.  When  sub-

merging test vessels, make sure that thermal fluid 

level will not go over the edges of the tank. Put a 

hose  on  valve  socket to  drain  the  excess  of  ther-

mal fluid. 

When  operating  instrument  with  thermal 

fluid  temperature  close  to  ambient,  it  might  be 

necessary  to  provide  the  cooling  by  means  of 

built-in  coil.  In  order to do  that,  connect  thermo-

stat  to  tap  water  supply  with  hoses,  attached  to 

the coil connectors. The flow of the cooling water 
must be even and slightly weak. The cooling is not 

necessary  if  thermal  fluid  temperature  is  at  least 

15 

°С

 higher than ambient temperature. 

To operate the instrument, read the “M03

Heating Immersion Circulator. Operating manual”.

 

 

 


 

 

 
 

 
 

 

 

 
 

 
 

 
 

Summary of Contents for VT-R-01

Page 1: ...HEATING CONSTANT TEMPERATURE BATHS VT R 01 VT R 03 Operating manual ...

Page 2: ...ure baths Operating manual 2 TABLE OF CONTENTS Introduction 3 Intended use 3 Appearance and parts names 3 Environmental Conditions 4 Safety Recommendations 4 Using The Heating constant temperature Baths 5 Preparation 5 GENERAL SPECIFICATIONS 6 ...

Page 3: ...standard test method for vapor pressure of petroleum products Reid Method according to ASTM D323 and ASTM D1267 Appearance and parts names Heating constant temperature baths consist of M03M heating immersion circulator 1 and a bath tank 2 The bath tank includes built in cooling coil 4 and drain valve 5 When running tests install Reid vapor pressure vessels with pressure gauges 3 Adapter 6 is inten...

Page 4: ...rried out by means of heat exchange with environment or cooling liquid passed through built in cooling coil 3 Environmental Conditions Indoor use only Ambient temperature 10 35 С Air humidity max relative humidity 80 for temperatures up to 31 С Max mains fluctuation of 10 are permissible Safety Recommendations Avoid strikes to the housing vibrations damage to the operating element panel keypad dis...

Page 5: ... to the power source Attach a pressure gauge to a Reid vapor pressure vessel Insert the vessel into adapter 6 and slide into the slot Fill up the bath tank with thermal fluid through aperture under bath cover 5 When sub merging test vessels make sure that thermal fluid level will not go over the edges of the tank Put a hose on valve socket to drain the excess of ther mal fluid When operating instr...

Page 6: ...ability 0 1 С Temperature uniformity 0 1 С Digital setting accuracy 0 5 С Digital setting repeatability 0 05 С Heating capacity 2000 W Bath volume VT R 01 VT R 03 30 Liters 40 Liters Dimensions W D H VT R 01 VT R 03 335 280 780 mm 545 250 780 mm Bath opening VT R 01 VT R 03 120 210 mm 290 160 mm Bath depth 500 mm Weight VT R 01 VT R 03 22 kg 32 kg Power supply 230 V 50 Hz 2 2 kW Warranty 2 years ...