TEREN H17P Operation Manual Download Page 1

H17P Flush Mount T/RH Display Unit/Transmitter -Operation Manual                                                          Edition:A/1

 

1

 

Flush Mount T/RH Display Unit/Transmitter 

 

 

 

 

Applications & Features 

 

Apply  for  flush  mount,  measure  and  display  environment 
T/RH  such  as  cleanrooms  in  electronics  or  pharmaceutical 
plants, hospital and various laboratories, etc. 

 

316L front panel, PMMA window, flat surface, no dust stays, 
good for all detergents, sanitizers 

and

 bactericides 

 

With  monochrome  LCD  display,  specially  designed  for  a 
variety  of  imported  well-known  brands  (E+E  and  Rotronic) 
probe products 

 

High 

accuracy, 

100% 

field 

interchangeable 

without 

re-calibration 

 

2  channels  4~20mA  output,  with  over  voltage  and  reverse 
polarity protection 

 

Strong  EMI  capability,  meet  the  requirements  of  EN61326-1 
industrial equipment 

 

Compatible  to  any  DDC/PLC/SCADA  or  other  data  collect 
and control systems 

Specifications 

T/RH  remote  probe: 

(

Refer to the operation manual from the original 

manufacturers for details

)

   

Main unit and display

4 bits LCD (with backlight), alternately 

display  T/RH,  display  area  47x27mm,  with  figure  16mm  and 
unit 2mm

 

Display panel material:

 PMMA 

Resolution:

 0.1°C(or °F, switchable), 0.1%RH 

Update time: 

<1s 

Power Supply:

16~28VAC/19~35VDC 

Consumption: 

1.5 VA 

Analog outputs:

2×4~20mA(3/4 wires), R

L

<500

Ω;

 

accuracy as 

low as 0.1%FS(T)/ 0.3%FS(RH) 

 

Terminals: 

max Ø1.5mm

2

 

Work Environment: 

0~50°C, 0~95%RH (no cond.) 

Storage Environment: 

-10~70°C 

Front panel material: 

316L stainless steel, 1.5mm thick 

Back housing parts:

 fire retardant ABS+PC(UL94V-0) 

Protection: 

front panel IP65   

Weight:

 

330g

 

Approval: 

CE, meet EN61326-1 for industrial equipment 

Note:  For  completeness,  the  above  specifications  include  some  of  the 
probe's function and technical parameters. The products are combination of 
a main unit and a separated probe.   

Models for the main unit 

(The product has two parts: the 

main unit and the probe. This is only for the main unit)

 

Model 

H17P 

Flush Mount T/RH Display Unit/Transmitter (basic model) 

Models  for  the  probe 

(The product has two parts: the main 

unit and the probe. This is only for the probe)   

Model 

Temp. and humidity probe, brand and accuracy

 

EE07-PFT3

 

E+E

, 2%RH@0~90%RH/0.1°C@20°C 

HC2A-S1.5-OEM

 

Rotronic

, 1.5%RH/0.3°C @10~30°C, 23°C, 10/35/80%RH 

HC2A-S

 

Rotronic

, 0.8%RH/0.1°C @10~30°C, 23°C, 10/35/80%RH 

HC2A-S-HH 

Rotronic

, 0.8%RH/0.1°C @10~30°C, 23°C, 10/35/80%RH, 

Hydrogen peroxide probe

 

1. The cable should be ordered separately, with length of 2m, 5m or 10m. The models 
for E+E are E2, E5 and E10 respectively, and for Rotronic are R2, R5 and R10. 
2.  The  main  unit,  probe  and  cable  should  be  selected  and  ordered  respectively.  For 
example:  the  main  unit  is  H17P,  the  probe  is  EE07PFT3  and  the  cable  is  E5.  The 
factory  will  deliver  the  three  parts  separately,  and  customer  will  install  and  connect 
them on site. Please refer to the operation manual for details. 
3.  In  the  environment  where  the  pharmaceutical  factory  uses  hydrogen  peroxide 
disinfection,  the  hydrogen  peroxide  filter HA010115  is  optional  for  the  probe  of  E+E, 
and  the  hydrogen  peroxide  probe  HC2A-S-HH  is  optional  for  Rotronic  to  resist  the 
sensor erosion during the hydrogen peroxide disinfection process. 
 

Electrical Connection

 

Due to the different selections, the terminals and wiring will be different, and the specific wiring should be wired according to the 
corresponding wiring diagram of the product model. 

H17P mainframe with EE07 probe wiring diagram:                  Brand

 

E+E,Model

 

EE07

 

probe cable wiring diagram:

 

Probe model

 

EE07

 

Cable

 

Cable color

 

white

 

brown

 

black

 

blue

 

argent

 

Electrical

 

signals

 

5V

 

GND

 

SCL

 

SDA

 

shield

 

H17P mainframe with HC2A probe wiring diagram:                  Brand Rotronic,Model HC2A Probe Cable Wiring Diagram: 

Probe model

 

HC2A

 

Cable

 

Cable color

 

grey

 

green

 

blue

 

red

 

argent

 

Electrical

 

signals

 

GND

 

5V

 

TX

 

RX

 

shield

 

 

 

EE07-PFT3 

  V-      V+

 OUT-H  OUT-T  GND

GND      5V       TX      RX

  1       2       3       4        5       6       7        8       9       10      11

A

A

POWER  SUPPLY

24VAC/DC

H

T

GND

Output

(4-20mA)

ON

 1      2      3     4     5

SDA   SCL

   EE07

  V-      V+

 OUT-H  OUT-T  GND

GND      5V       TX      RX

  1       2       3       4        5       6       7        8       9       10      11

HC2A

A

A

POWER  SUPPLY

24VAC/DC

H

T

GND

Output

(4-20mA)

ON

 1      2      3     4     5

SDA   SCL