background image

www.

grupotemper.com

 

 

 

 
 

Pinza amperimétrica TRMS 

Pinça amperimetrica TRMS 

Clamp meter TRMS 

Pince amperimètrique TRMS 

 
 
 
 

v1.0 

KPA-11T 

0767498 

 

Summary of Contents for 0767498

Page 1: ...www grupotemper com Pinza amperim trica TRMS Pin a amperimetrica TRMS Clamp meter TRMS Pince amperim trique TRMS v1 0 KPA 11T 0767498...

Page 2: ...do Temperatura Hz Apagado autom tico Retenci n de datos Retenci n de picos Pantalla LCD con retroiluminaci n 2 SEGURIDAD 2 1 S mbolos internacionales de seguridad Este s mbolo junto a un terminal indi...

Page 3: ...s volta jes conllevan un alto riesgo de descarga el ctrica Descargue los capacitores y retire la alimentaci n del dispositivo bajo prueba antes de realizar pruebas de diodo resistencia o continuidad L...

Page 4: ...I N 1 Pinza amperim trica 2 Gatillo 3 Selector de funci n 4 Bot n HOLD 5 Bot n PEAK retroiluminaci n 6 Bot n MODE 7 Bot n RANGE 8 Bot n REL Hz 9 Pantalla LCD 10 Conector de entrada COM 11 Conector V C...

Page 5: ...C a cero PEAK Captura de picos AUTO Modo de rango autom tico DC AC Corriente continua corriente alterna Bater a baja mV o V Milivoltios o voltios voltaje Ohmios resistencia A Amperios corriente F Fara...

Page 6: ...specificados desde el 5 al 100 del rango Funci n Rango Precisi n de lectura d gitos Resoluci n Corriente de CA True RMS 400 A 3 3 d g 100 mA 2000 A 3 5 6 d g 1 A Protecci n contra sobre pasamiento de...

Page 7: ...to de rango 600 V rms Funci n Rango Precisi n de lectura d gitos Resoluci n Pico de corriente de CA 400 A 4 5 d g 100 mA 2000 A 4 5 d g 1 A Respuesta de frecuencia 50 Hz 60 Hz Funci n Rango Precisi n...

Page 8: ...pico Captura picos 1 ms Sensor de temperatura Termopar tipo K Impedancia de entrada 10 M V CC y V CA Temperatura de funcionamiento De 5 C a 40 C Temperatura de almacenamiento De 20 C a 60 C Humedad d...

Page 9: ...o Coloque el selector de funci n en la posici n de CA CC Conecte los cables de prueba en paralelo al circuito bajo prueba Lea la medici n del voltaje en la pantalla LCD 7 3 Medici n de resistencia Ins...

Page 10: ...antes de cambiar a otra funci n de medici n 7 7 Medici n de continuidad Inserte el cable de prueba negro en el terminal negativo COM y el rojo en el terminal positivo Coloque el selector de funci n e...

Page 11: ...ici n A o V 2 Use el bot n MODE para seleccionar AC o DC 3 Espere unos instantes hasta que la pantalla se estabilice 4 Pulse el bot n PEAK se mostrar la leyenda PEAK en la pantalla 5 La pantalla se ac...

Page 12: ...abrasivos ni disolventes Si no va a usar el medidor durante 60 d as o m s retire la bater a y almacene el medidor y la ba ter a por separado 8 2 Sustituci n de la bater a Con un destornillador Phillip...

Page 13: ...mento autom tico Manuten o de dados Manuten o de pico Visor LCD retroiluminado 2 SEGURAN A 2 1 S mbolos Internacionais de Seguran a Este s mbolo adjacente a um terminal indica que em utiliza o normal...

Page 14: ...forem superiores a 25VAC rms ou 35VDC Essas tens es s o consideradas como um perigo de choque Descarregue sempre os condensadores e remova a energia do dispositivo em teste antes de realizar testes de...

Page 15: ...ente 2 Gatilho da garra 3 Interruptor da fun o rotativa 4 Bot o de reten o de dados 5 Bot o de PICO Retroilumina o 6 Bot o de sele o de modo 7 Bot o de alcance 8 Bot o de relativa Hz 9 Visor LCD 10 Co...

Page 16: ...Pico Captar pico AUTO Modo de intervalo autom tico DC AC Corrente cont nua e corrente alterna Bateria fraca mV ou V Mili volts ou volts voltagem Ohms Resist ncia A Amperes Corrente F Farad Capacit nc...

Page 17: ...os de 5 do intervalo a 100 do intervalo Fun o Intervalo Precis o de leitura d gitos Resolu o Corrente AC Valor eficaz verdadeiro 400A 3 0 3d 100mV 2000A 3 5 6d 1A Prote o acima do intervalo Entrada m...

Page 18: ...atura 20 a 1000 C 2 0 5 C 0 1 C 4 a 1832 F 3 0 9 F 0 1 F Sensor Reprodu o Tipo K em formato de banana Prote o acima do intervalo 600V valor eficaz verdadeiro Fun o Intervalo Precis o de leitura d gito...

Page 19: ...r de temperatura Termopar de tipo K Imped ncia de entrada 10M VDC e VAC Temperatura de funcionamento 41 F a 104 F 5 C a 40 C Temperatura de armazenamento 4 F a 140 F 20 C a 60 C Humidade de funcioname...

Page 20: ...terminal positivo Coloque o interruptor de fun o na posi o AC DC Ligue os terminais de teste em paralelo ao circuito sobre teste Leia de medi o de tens o no visor LDC 7 3 Medi es de resist ncia Insira...

Page 21: ...do para selecionar F ou C AVISO Para evitar choques el tricos assegure se de que a sonda termopar foi removida antes de mudar para outra fun o de medi o 7 7 Medi es de Continuidade Insira o terminal d...

Page 22: ...de Pico capta a tens o ou corrente AC ou DC de pico O medidor pode captar picos negativos ou positivos t o r pidos quando 1 milissegundo de dura o Coloque o interruptor de fun o na posi o A ou V Use o...

Page 23: ...o h mido e detergente suave n o use produtos abrasivos ou solventes Se o medidor n o for usado por mais de 60 dias remova a bateria e guarde a separadamente 8 2 Substitui o da Bateria Remova o parafus...

Page 24: ...ce Frequency Continuity Diode Temperature Hz Auto Power OFF Data Hold Peak Hold Backlight LCD display 2 SAFETY 2 1 International Safety Symbols This symbol adjacent to a terminal indicates that under...

Page 25: ...5VDC These voltages are considered a shock hazard Always discharge capacitors and remove power from the device under test before performing diode re sistance or continuity tests Voltage checks on elec...

Page 26: ...3 DESCRIPTION 1 Clamp Meter 2 Trigger 3 Function selector 4 HOLD button 5 PEAK button backlight 6 MODE button 7 RANGE button 8 REL Hz button 9 LCD screen 10 COM input connector 11 Connector V CAP TEM...

Page 27: ...ue A CC to zero PEAK Peak capture AUTO Automatic range mode DC AC Direct current alternating current Low battery mV o V Millivolts or volts voltage Ohms resistance A Amperes current F Farad capacitanc...

Page 28: ...C current Rang are specified from 5 of rang to 100 of rang Function Range Accuracy of reading digits Resolution AC Current TRMS 400 A 3 3 dig 100 mV 2000 A 3 5 6 dig 1 A Over rang protection Maximum i...

Page 29: ...rang protection 600 V rms Function Range Accuracy of reading digits Resolution AC current peak 400 A 4 5 dig 100 mA 2000 A 4 5 dig 1 A Frequency behaviour 50 Hz 60 Hz Function Range Accuracy of readin...

Page 30: ...nal Peak value Capture peaks 1 ms Temperature sensor Thermocouple type K Input impedance 10 M V DC and V AC Operating temperature From 5 C to 40 C Storage temperature 20 C to 60 C Operating humidity M...

Page 31: ...nto the negative COM terminal and the red test lead into the Set the function switch to the AC DC position Connect the test leads in parallel to the circuit under test Read the voltage measurement on...

Page 32: ...dicate the proper decimal point and value Use the MODE button to select C or F WARNING to avoid electric shock be sure the thermocouple probe has been removed before changing to another measurement fu...

Page 33: ...ction captures the peak AC or DC voltage or current The meter can capture negative or pos itive peaks as fast as 1 millisecond in duration Turn the function switch to the A or V position Use the MODE...

Page 34: ...clamp cloth and mild detergent do not use abrasives or solvents If the meter is not to be used for 60 days or more remove the battery and store it separately 8 2 Battery Replacement Remove the Philip...

Page 35: ...t auto de l alimentation Maintien des donn es Maintien de cr te Affichage de r tro clairage LCD 2 SECURITE 2 1 Symboles de s curit internationale Ce symbole adjacent un terminal indique l utilisateur...

Page 36: ...ent 25VAC rms ou 35VDC Ces tensions sont consid r es comme risque d lectrocution Toujours d charger les capaciteurs et retirer l alimentation de l appareil sous test avant d effectuer les tests de dio...

Page 37: ...cheur de pince 3 Commutateur de fonction tournante 4 Bouton de maintien des donn es 5 Bouton de CRETE Bouton de r tro clairage 6 Bouton de s lection du MODE 7 Bouton de plage 8 Bouton relatif Hz 9 Aff...

Page 38: ...res de mesure REL DCA Z ro CRETE Capture de cr te AUTO Mode de plage auto DC AC Courant direct courant alternatif Batterie faible mV ou V Millivolts ou Vote Tension Ohms R sistance A Amp res Courant F...

Page 39: ...outes les plages de tension AC sont sp cifi es dans les plages de 5 100 Fonction Plage Fiabilit de lecture chiffres R solution Courant AC T RMS 400A 3 0 3 chiffres 100mV 2000A 3 5 6 chiffres 1A Protec...

Page 40: ...tion Temp rature 20 1000 C 2 0 5 C 0 1 C 4 1832 F 3 0 9 F 0 1 F Capteur Lecture de banane de type K Protection hors plage 600V rms Fonction Plage Fiabilit de lecture chiffres R solution Cr te du coura...

Page 41: ...teur de temp rature Thermocouple de type K Entr e d imp dance 10M VDC et VAC Temp rature de fonctionnement 41 F 104 F 5 C 40 C Temp rature de stockage 4 F 140 F 20 C 60 C Humidit de fonctionnement Max...

Page 42: ...mutateur de fonction la position AC DC Connecter les t tes de test en parall le au circuit en cours de test Lire la mesure de tension sur l affichage LCD 7 3 Mesures de r sistance Ins rer la t te de t...

Page 43: ...appropri s Utiliser le bouton du MODE pour s lectionner F ou C AVERTISSEMENT Pour viter l lectrocution s assurer que la sonde du thermocouple a t retir e avant de changer une autre fonction de mesure...

Page 44: ...a cr te de tension ou du courant AC ou DC Le compteur pourrait capturer les cr tes positives et n gatives en moins de 1 milliseconde R gler le commutateur de fonction la position V ou A Utiliser le bo...

Page 45: ...t instrument Si le compteur ne sera pas utilis pendant 60 jours ou plus retirer la batterie et la stocker s par ment 8 2 Remplacement de la batterie Retirer la vis molet e Phillips s curisant le couve...

Page 46: ...tout default de fabrication Pour valider cette garantie il est essential d avoir la facture d achat EN T E I Guarantees this product for 2 years against any manufacturing defect To make this guarantee...

Reviews: