Instruction Manual
MX7M Active Gauge
Pr essr e
(
Tor r
) =
10
3
.
075
2
.
5
−
7
=
1
.
7
×
10
−
6
Tor r
To select the Logarithmic output, navigate to the Analog Output screen, press ENT, and then use
the UP or DOWN arrow to select Logarithmic option. Press ENT to save the change. Press the
DOWN arrow to navigate to the 10 V exponent selection screen. Press ENT to unlock and then
the UP or DOWN arrow to select the value. Press ENT to save the value. Press the DOWN arrow
to navigate to the 0 V exponent selection screen. Press ENT to unlock the screen and the UP or
DOWN arrow to select the 0 V value.
Figure 3: Graph of MX7M Sample Analog Output Functions
email: [email protected]
tel: +1 215 947 2500
web: frederickscompany.com
mx7m_im rev C