Instruction Manual
MX7M Active Gauge
3.6 RS-485 and USB Communications Error Codes
Table 12: RS-485 and USB Communications Error Codes
Error Codes
Description
0N0001<cr><lf>
Command error: an invalid character was sent for the
command or an invalid number after the character.
0N0002<cr><lf>
Units error: an invalid parameter was sent after a com-
mand specifier
Table 13: RS-485 and USB Status Commands
Command
Response
Description
S1
ppsee
Read pressure data where ppsee is:
pp = mantissa
s = sign
ee = exponent
When in the “Starting” state, unit returns “Starting”
When in the “Low” state, unit returns “Low”
When high voltage is off, unit returns "0"
S2
yymmdd
Read firmware version, where:
yy = last two digits of year
mm = month
dd = day
3.7 Firmware Updates
The MX7M has the ability to update the firmware when in the field over the USB communications. For
more information on updating the firmware, please contact
email: [email protected]
tel: +1 215 947 2500
web: frederickscompany.com
mx7m_im rev C
23 of 23