Telemark TT Cheetah Series Power Supply Instruction Manual
18
Pin#
Name
Description
1
Phase 1
AC Phase 1
3
GND / PE
AC GND / PE
4
Neutral
AC Neutral,
on 400VAC units
only
5
Phase 2
AC Phase 2
6
Phase 3
AC Phase 3
AC Line Input TT-10/15/20
(Terminal Block, pin count from left)
#
Name
Description
1
P1
AC Phase 1
2
P2
AC Phase 2
3
P3
AC Phase 3
4
GND
AC GND / PE
5
N
AC Neutral,
on 400VAC units
only
J7
–
EXT EMO
–
not on TT-3/6
(Hirschmann 2-pin rectangular)
Pin#
Name
Description
1
External EMO
Interlock
Contact closure to pin 2 to enable HVPS,
open will disable HVPS
2
External EMO
Interlock
Contact closure to pin 1 to enable HVPS,
open will disable HVPS
Controller
J1
–
HVPS
–
HV CONTROL
>>> from J1 of HVPS
J2
–
HVPS
–
SOURCE CTRL
>>> from J2 of HVPS
J3/J4
–
SUB CONTROL #2/#3
>>> to J3/J4 of Sub Controller (see section 6)
J5
–
REMOTE USER I/O
(37pin Sub-D M)
Pin
#
Name
Input/
Output/
Power
Signal
Range
Function Description
1
User +24VDC
Power
+24VDC
Source for user +24VDC control
(200mA max output)
2
HV Ready/OFF
Indicator
Output
(O.C.)*
0 to +Vcc
Low = HV ready/off
High = HV not ready
3
Arc Indicator
Output
(O.C.)*
0 to +Vcc
Low = HV arcing/not regulating
High = HV regulating
4
Air/Timer Indicator
Output
(O.C.)*
0 to +Vcc
Low = Air Flow not OK
Low/High pulsing = Time Delay
High = Air Flow / Time Delay OK