BEAM CONTINUITY TEST
When prompted to do so, vigorously rub the selected
beam's face.
If a beam does not PASS the test, send any character to
the H-ADCP to automatically select the next beam.
Collecting Statistical Data...
52 48 50 43
Rub Beam 1 = PASS
Rub Beam 2 = PASS
Rub Beam 3 = PASS
Rub Beam 4 = PASS
This test must be run while the H-ADCP is in air.
PC2 – Display Heading, Pitch, Roll, and Orientation
Sending PC2 displays heading, pitch angle, roll angle, up/down orientation and attitude temperature in a
repeating loop at approximately 0.5-sec update rate. Press any key to exit this command and return to the
command prompt.
Press any key to quit sensor display ...
Heading Pitch Roll Up/Down Attitude Temp Ambient Temp Pressure
301.01
°
-7.42
°
-0.73
°
Up 24.35
°
C 22.97
°
C 0.0 kPa
300.87
°
-7.60
°
-0.95
°
Up 24.36
°
C 22.97
°
C 0.0 kPa
300.95
°
-7.60
°
-0.99
°
Up 24.37
°
C 22.97
°
C 0.0 kPa
300.71
°
-7.61
°
-0.96
°
Up 24.37
°
C 22.98
°
C 0.0 kPa
300.69
°
-7.61
°
-0.96
°
Up 24.35
°
C 22.98
°
C 0.0 kPa
300.76
°
-7.60
°
-0.98
°
Up 24.38
°
C 22.97
°
C 0.0 kPa
The PC2 heading shows the raw (magnetic north) heading only. The EB command (Heading Bias)
is not applied.
PD - Data Stream Select
Purpose:
Selects the type of ensemble output data structure.
Format:
PDn
Range
n = 0 to 18 (see description)
Default
PD0
Recommended Setting.
The default setting for this command is recommended for most applications.
Description:
PD selects the normal output data structure, a special application data structure, or a
fixed data set for transmission/display as the data ensemble (see Table 23).