Teledyne API – Model T400 O
3
Analyzer
083730400A DCN8130
Table 1-2. IZS Generator Specifications
with
Reference Feedback Option
PARAMETER
SPECIFICATION
Maximum Concentration
1.0 ppm
Minimum Concentration
0.050 ppm
Resolution
0.5 ppb
Repeatability (7 days)
1% of reading
Initial Accuracy
+/- 5% of target concentration
Response Time
< 5 min to 95%
Table 1-3. IZS Generator Specifications
without
Reference Feedback Option
PARAMETER
SPECIFICATION
Maximum Concentration
1.0 ppm
Minimum Concentration
0.050 ppm
Resolution
0.5 ppb
Repeatability (7 days)
2% of reading
Initial Accuracy
+/- 10% of target concentration
Response Time
< 5 min to 95%
1.2 EPA DESIGNATION
Teledyne API’s Model T400 photometric ozone analyzer is officially designated as
US EPA Federal Equivalent Method (FEM), Designation Number EQOA-0992-087. The
official “List of Designated Reference and Equivalent Methods” is published in the U.S.
http://www3.epa.gov/ttn/amtic/criteria.html
and specifies the
instrument’s settings and configurations.
1.3 SAFETY
IEC/EN 61010-1:2010 (3
rd
Edition), Safety requirements for electrical equipment for
measurement, control, and laboratory use.
CE:
2006/95/EC, Low-Voltage Directive
1.4 EMC
EN 61326-1 (IEC 61326-1), Class A Emissions/Industrial Immunity
FCC 47 CFR Part 15B, Class A Emissions
CE:
2004/108/EC, Electromagnetic Compatibility Directive
1.5 OTHER TYPE CERTIFICATIONS
EN 15267 Air Quality – Automated Measuring Systems
EN 14625 Ambient Air
Sira MC 050070/08
For additional certifications, please contactTechnical Support.