78
Figure 3-12 The main interface of the ExtI mode
The parameters for the ExtI mode include RangI and RangV.
Set Range
Current range: 5 A or 30 A
Voltage range: 36 V or 150 V
Set External Input Voltage
The sink current in ExtI mode can be controled by the external input
voltage which range is 0-10 V. The default unit of current is A
3.
Turn on the input channel
Press On/Off to turn on the channel input. At this time, the actual sink
voltage, current, power, resistance will be displayed on the main
interface.