T3CP100-2 Current Probe Manual

100 Amp, 2 MHz Current Probe

Summary of Contents for T3CP100-2

Page 1: ...T3CP100 2 Current Probe Manual 100 Amp 2 MHz Current Probe ...

Page 2: ...This page is intentionally blank ...

Page 3: ...ions 7 Certifications 11 Safety Compliance 12 Environmental Compliance 12 RoHS 13 Warranty 14 2021 Teledyne LeCroy Inc All rights reserved Teledyne Test Tools is a trademark of Teledyne LeCroy Inc Other product or brand names are trademarks or requested trademarks of their respective holders Information in this publication supersedes all earlier versions Specifications are subject to change withou...

Page 4: ... be applied around or removed from insulated limited energy circuit conductors that are not directly connected to the mains Do not overload observe all ratings To avoid electric shock or fire do not connect the current probe to any wire that carries voltages or currents that exceed the ratings of the probe Connect and disconnect properly Connect the probe to the test instrument before connecting t...

Page 5: ...tions may result in electric shock Keep product surfaces clean and dry Comply with the maximum input current vs frequency derating when measuring current that includes a high frequency component Using the probe at high frequencies or in strong magnetic fields may cause the device to become abnormally hot resulting in fire equipment damage or burns Do not operate with suspected failures Before each...

Page 6: ... Teledyne Test Tools new T3CP100 2 current probe is a 2MHz bandwidth active AC DC coupled probe featuring 2 ranges of 100 A Peak 70 A rms 10A Peak 7 A rms fast and accurate waveform capture risetime of 175 ns and low test circuit loading This probe can be used with any oscilloscope having a 1MΩ BNC input Applications Power design and power component measurements Consumer electronics and household ...

Page 7: ...rnal power supply supplied is connected to the T3CP100 2 current probe High Low Range Indicators During use of the probe one of the high low indicators will always be lit indicating the current measurement range selected The range can be changed by using the High Low Range Selector High Low Range Selector This button allows the high or low range of the probe to be selected T3CP100 2 Product Descri...

Page 8: ...e external power supply 1m Interconnect Cable The 1m current probe head control unit interconnect cable is permanently attached between the probe head and the control unit Probe Head The current probe measurement head uses a spring loaded lever to open and close the jaw The jaw can clamp around cables up to 12mm in diameter The measurement cable DUT should always be insulated Do not clamp around u...

Page 9: ...move the probe measurement jaw from the circuit under test Since the probe has a 2 MHz bandwidth it is recommended to turn on a bandwidth limit filter on the Oscilloscope if it has one Safe Probing This device is designed to comply with Safety Standards and has been thoroughly tested for safety prior to shipment However mishandling during use could result in injury or death as well as damage to th...

Page 10: ...BNC to BNC probe output connecting cable to connect the T3CP100 2 BNC output to an Oscilloscope input Wall socket PSU The wall socket PSU can power the probe in place of the batteries Maintenance Keep the probe dry and clean Clean the probe with a dry cloth do not use liquid cleaning agents After use put the probe back in its case and store in a clean and dry place out of the sun To maintain the w...

Page 11: ... I 7 07A 70 7A 600V 600V RMS CAT II 7 07A 70 7A 600V 600V RMS CAT III 7 07A 70 7A 300V 300V Voltage And Current Rating Table Safety Ratings Probe Designation1 Type D Measurement Category 2 O Other Safe Voltage Rating 3 Use only on fully insulated conductors Pollution Degree 4 2 Definitions per IEC EN 61010 2 032 2019 1 Type D current probe is intended to be applied around or removed from insulated...

Page 12: ...Page 9 Figure 2 Maximum Current Versus Frequency Figure 3 Typical DC Linearity High Range Figure 1 Bandwidth ...

Page 13: ...3mm x 35mm Operating Altitude 0 2000 meters Maximum Measurement Conductor Diameter 12mm Current Clamp To Control Unit Cable Length 1 meter Output BNC to BNC Cable Length 1 meter Weight 223 grams excluding 9V Battery Operating Temperature 00 C 500 C Storage Temperature 200 C 800 C Operating Relative Humidity 00 C 400 C 95 RH 400 C 500 C 45 RH Pollution Degree Level 2 ...

Page 14: ...ed Emissions Group 1 Class A 2 3 ELECTROMAGNETIC IMMUNITY EN 61000 4 2 2009 Electrostatic Discharge 4 kV contact 8 kV air 4 kV vertical horizontal coupling planes 4 EN 61000 4 3 A2 2010 RF Radiated Electromagnetic Field 3 V m 80 1000 MHz 3 V m 1400 MHz 2 GHz 1 V m 2 GHz 2 7 GHz EN 61000 4 8 2010 Power Frequency Magnetic Field 3 A m 50 Hz 3 A m 60 Hz 1 To ensure compliance with the applicable EMCst...

Page 15: ... for Product Safety Compliance was demonstrated to the following specifications as listed in the Official Journal of the European Communities IEC EN 61010 1 2010 A1 2019 Safety requirements for electrical equipment for measurement control and laboratory use Part 1 General requirements IEC EN 61010 2 032 2019 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use Pa...

Page 16: ...ces contained in the instrument are disclosed in accordance with the standards SJ T 11364 2014 Marking for the restricted use of hazardous substances in electronic and electrical products and GB T 26572 2011 Requirements on concentration limits for certain restricted substances in electrical and electronic products The instrument is marked with an appropriate Environmental Friendly Use Period EFUP...

Page 17: ...ve due to workmanship or materials and the defect is not caused by misuse neglect accident abnormal conditions of operation or damage resulting from attempted repair or modifications by a non authorized service facility The customer will be responsible for the transportation and insurance charges for the return of products to the service facility Teledyne LeCroy will return all products under warr...

Page 18: ...This page is intentionally blank ...

Page 19: ...This page is intentionally blank ...

Page 20: ... reduce time to market Designers engineers and educators rely on Teledyne Test Tools solutions to meet their most challenging needsfor testing education and electronics validation Location and Facilities Headquartered in Chestnut Ridge New York Teledyne Test Tools and Teledyne LeCroyhas sales service and development subsidiaries in the USand throughout Europeand Asia Teledyne Test Tools and Teledy...

Reviews: