Operating Instructions
Teledyne API Model T360/T360M Operation Manual
72
To view and clear warning messages:
SAMPLE
SYSTEM RESET CO2 = 0.00
TEST
CAL
MSG
CLR
SETUP
Press
CLR
to clear the current
message.
If more than one warning is active, the
next message will take its place
Once the last warning has been
cleared, the analyzer returns to
SAMPLE
mode
SAMPLE
RANGE=500.000 PPM
CO2 = 0.00
< TST TST > CAL
MSG
CLR
SETUP
SAMPLE
SYSTEM RESET
CO2 = 0.00
TEST CAL
MSG
CLR
SETUP
TEST
deactivates warning
messages
MSG
displays warning messages.
NOTE:
If the warning message persists
after several attempts to clear it,
the message may indicate a
real problem and not an artifact
of the warm-up period
Figure 4-3:
Viewing and Clearing T360 WARNING Messages
4.3. CALIBRATION MODE
Pressing the
CAL
button switches the T360 into multi-point calibration mode. In
this mode, the user can calibrate the instrument or check the instruments
calibration with the use of calibrated zero or span gases.
If the instrument includes either the zero/span valve option or IZS option, the
display will also include
CALZ
and
CALS
buttons. Pressing either of these
buttons also puts the instrument into multipoint calibration mode.
CALZ
is used to initiate a calibration of the zero point.
CALS
is used to calibrate the span point of the analyzer. It is recommended
that this span calibration is performed at 90% of full scale of the analyzer’s
currently selected reporting range.
Because of their critical importance and complexity, calibration operations are
described in detail in Section 5 of this manual. For more information concerning
the valve options, see Section 3.6.
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...