Getting Started
Teledyne API Model T360/T360M Operation Manual
58
3.8. INITIAL CALIBRATION PROCEDURE
The next task is to calibrate the analyzer.
To perform the following calibration you must have sources for zero air and span
gas available for input into the sample port on the back of the analyzer. See
Section 3.6 for instructions for connecting these gas sources.
While it is possible to perform this procedure with any range setting we
recommend that you perform this initial checkout using the 500 ppm range.
The following procedure assumes that the instrument does not have any of
the available Zero/Span Valve Options installed. See Section 5.4 for
instructions for calibrating instruments with Z/S valve options.
1. Set the Analog Output Range:
SETUP X.X
CFG DAS
RNGE
PASS CLK MORE
EXIT
SAMPLE RANGE = 500.000 PPM
CO2
=X.XXX
< TST TST > CAL
SETUP
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
Press this button to set
the analyzer for
SNGL
DUAL
or
AUTO
ranges
SETUP X.X
RANGE: 500.000 CONC
0 0
5
0 0 .
0
ENTR EXIT
To change the value of the
reporting range span, enter the
number by pressing the key under
each digit until the expected value
appears.
Press this button to select the
concentration units of measure:
PPB, PPM, UGM, MGM
EXIT
ignores the new setting and
returns to the
RANGE CONTROL
MENU
.
ENTR
accepts the new setting and
returns to the
RANGE CONTROL MENU
.
SETUP X.X
RANGE:
500.000 Conc
0 0 0 5 0 .0
ENTR
EXIT
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...