Getting Started
Teledyne API Model T360/T360M Operation Manual
52
Figure 3-19:
Internal Pneumatic Flow – Ambient Zero/Pressurized Span Valves
The following table describes the state of each valve during the analyzer’s
various operational modes.
Table 3-10:
Ambient Zero/Pressurized Span Valve Operating States
MODE VALVE
CONDITION
Sample/Cal
Open to SAMPLE inlet
Zero/Span
Open to IZS inlet
SAMPLE
(Normal State)
Shutoff Valve
Closed
Sample/Cal
Open to ZERO/SPAN valve
Zero/Span
Open to IZS inlet
ZERO CAL
Shutoff Valve
Closed
Sample/Cal
Open to ZERO/SPAN valve
Zero/Span
Open to SHUTOFF valve
SPAN CAL
Shutoff Valve
Open to PRESSURE SPAN Inlet
The minimum span gas flow rate required for this option is 800 cm
3
/min.
The state of the zero/span valves can also be controlled:
Manually from the analyzer’s front panel by using the
SIGNAL I/O
controls
located under the
DIAG
Menu (Section 4.13.2)
By activating the instrument’s AutoCal feature (Section 5.6)
Remotely by using the external digital control inputs (Section 4.15.1.2)
Remotely through the RS-232/485 serial I/O ports (see Appendix A-6 for the
appropriate commands).
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...