Teledyne API Model T360/T360M Operation Manual
Operating Instructions
151
4.14. SETUP
MORE
ALRM: USING THE GAS
CONCENTRATION ALARMS
The Model T360 includes two CO
2
concentration alarms. Each alarm has a user
settable limit, and is associated with an opto-isolated TTL relay accessible via the
status output connector on the instrument’s back panel (see Section 4.15.1.1). If
the CO
2
concentration measured by the instrument rises above that limit, the
alarm‘s status output relay is closed.
The default settings for
ALM1
and
ALM2
are:
Table 4-22: CO2 Concentration Alarm Default Settings
ALARM STATUS
LIMIT SET POINT
1
ALM1
Disabled 100
ppm
ALM2
Disabled 300
ppm
1
Set points listed are for PPM. Should the reporting range units of measure be changed
(see Section 4.8.2) the analyzer will automatically scale the set points to match the new
range unit setting.
Note
To prevent the concentration alarms from activating during span calibration operations make sure
to press CAL or CALS button prior to introducing span gas into the analyzer.
4.14.1. Setting the Concentration Alarm Limits
To enable either of the CO
2
concentration alarms and set the Limit points, press:
ENTR
accepts the new
settings
EXIT
ignores the new
settings
SETUP X.
ALARM 1 LIMIT: 200.00 PPM
0 1 0 0 .0 0
ENTR EXIT
SETUP X.X
ALARM MENU
ALM1 ALM2
EXIT
SETUP X.
ALARM 1 LIMIT: OFF
OFF
ENTR EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG
ALRM
EXIT
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
Toggle these buttons to
scroll through the
available character set:
0-9
SETUP X.
ALARM 1 LIMIT: ON
ON
ENTR EXIT
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...