Operating Instructions
Teledyne API Model T360/T360M Operation Manual
146
4.13.5. Electric Test
The electric test function substitutes simulated signals for
CO
2
MEAS
and
CO
2
REF
, generated by circuitry on the sync/demod board, for the output of the IR
photo-detector. While in this mode the user can also view the same test functions
viewable from the main
SAMPLE
display. When the test is running, the
concentration reported on the front panel display should be
40.0
ppm.
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
ALRM
EXIT
Press
<TST TST>
to view Test Functions
NOTE:
CO
2
MEAS
and
CO
2
REF
will be artificially
altered
to enforce a CO
2
reading of
40.0
ppm.
All other Test Functions will report the correct operational
value
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
Exit returns
to the
DIAG
Menu
Repeat Pressing
NEXT
unti . . .
DIAG
ELECTRIC TEST
PREV NEXT
ENTR
EXIT
DIAG ELEC RANGE=50.000 PPM
CO2= 40.0
<TST TST>
EXIT
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
ENTER DIAG PASS: 818
8 1 8
ENTR
EXIT
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...