Getting Started
Teledyne API – Model T300/T300M CO Analyzer
74
3.3.2.10.
PNEUMATIC LAYOUT FOR ZERO SCRUBBER/ AMBIENT SPAN OPTION
INSTRUMENT CHASSIS
Flow / Pressure
Sensor PCA
SAMPLE
PRESSURE
SENSOR
FLOW
SENSOR
Sample Gas
Flow Control
SAMPLE
GAS INLET
EXHAUST
GAS OUTLET
Particulate
Filter
PUMP
ZERO AIR
INLET
SPAN1
INLET
VENT SPAN
OUTLET
Sample / Cal
Valve
Zero / Span
Valve
GFC Wheel
Housing
GFC Motor Heat
Sync
SAMPLE CHAMBER
ZERO Air
Scrubber
Zero
Gas
In
Sample
Gas
In
Span
Gas
In
Figure 3-28:
Internal Pneumatic Flow OPT 50H – Zero Scrubber/Ambient Span
Table 3-11: Zero/Span Valve Operating States for Option 50H
MODE
VALVE
CONDITION
SAMPLE
(Normal
State)
Sample/Cal
Open to SAMPLE inlet
Zero/Span
Open to ZERO AIR scrubber
ZERO CAL
Sample/Cal
Open to ZERO/SPAN valve
Zero/Span
Open to ZERO AIR scrubber
SPAN CAL
Sample/Cal
Open to ZERO/SPAN valve
Zero/Span
Open to PRESSURE SPAN inlet
3.3.2.11.
CALIBRATION GASES
Zero air and span gas are required for accurate calibration.
Z
ERO
A
IR
Zero air is a gas that is similar in chemical composition to the earth’s atmosphere but
scrubbed of all components that might affect the analyzer’s readings, in this case CO
and water vapor. If your analyzer is equipped with an Internal Zero Span (IZS) or an
external zero air scrubber option, it is capable of creating zero air.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...