Teledyne API – Model T300/T300M CO Analyzer
Getting Started
65
3.3.2.2.
PNEUMATIC LAYOUT FOR BASIC CONFIGURATION
INSTRUMENT CHASSIS
Flow / Pressure
Sensor PCA
SAMPLE
PRESSURE
SENSOR
FLOW
SENSOR
Sample Gas
Flow Control
SAMPLE GAS
INLET
EXHAUST
GAS OUTLET
Particulate Filter
PUMP
GFC Wheel
Housing
GFC Motor Heat Sync
SAMPLE CHAMBER
Figure 3-20:
T300/T300M Internal Gas Flow (Basic Configuration)
3.3.2.3.
PNEUMATIC CONNECTIONS FOR AMBIENT ZERO/AMBIENT SPAN VALVE OPTION
This valve option is intended for applications where:
•
Zero air is supplied by a zero air generator like the Teledyne API’s T701 and;
•
Span gas is supplied by Gas Dilution Calibrator like the Teledyne API’s T700.
Internal zero/span and sample/cal valves control the flow of gas through the instrument,
but because the generator and calibrator limit the flow of zero air and span gas, no
shutoff valves are required.
See Figure 3-4 for the location of gas inlets.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...