Teledyne API – Model T300/T300M CO Analyzer
Troubleshooting and Service
281
12.5.7.
SENSOR ASSEMBLY
12.5.7.1.
SYNC/DEMODULATOR ASSEMBLY
To verify that the Sync/Demodulator Assembly is working, follow the procedure below:
1. Verify that D1 and D2 are flashing.
•
If not, check the opto pickup assembly, Section 12.5.7.3 and the GFC Wheel
drive, Section 12.5.7.4.
•
If the wheel drive and opto pickup are working properly then verify that there is
2.4 ±0.1 VAC and 2.5 ±0.15 VDC between digital ground and TP 5 on the
sync/demod board. If not then check the wiring between the sync/demod and
opto pickup assembly (see interconnect drawing, P/N 04216). If good then the
sync/demod board is bad.
2. Verify that the IR source is operating, Section 12.5.7.5.
3. With the analyzer connected to zero air, measure between TP11 (measure) and
analog ground, and TP12 (reference) and analog ground.
•
If they are similar to values recorded on the factory data sheet then there is likely
a problem with the wiring or the A/D converter.
•
If they are not then either the sync demodulator board or the IR-photodetector are
bad. See Section 12.4.1.4 for problems with the IR-photodetector TEC drive.
12.5.7.2.
ELECTRICAL TEST
The electric test function substitutes simulated signals for CO MEAS and CO REF,
generated by circuitry on the sync/demod board, for the output of the IR photo-detector.
While in this mode the user can also view the same test functions viewable from the
main
SAMPLE
display. When the test is running, the concentration reported on the
front panel display should be 40.0 ppm. Also see Section 5.9.4.
SAMPLE
RANGE=50.0 PPM CO= XX.XX
<TST TST> CAL
SETUP
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
DIAG
SIGNAL I/O
PREV
NEXT
ENTR EXIT
SETUP X.X
ENTER PASSWORD
8
1
8
ENTR
EXIT
Continue pressing NEXT until ...
DIAG OPTIC
ELECTRICAL TEST
PREV NEXT
ENTR
EXIT
DIAG ELEC
RANGE=50.0 PPM CO= 40
<TST TST> CAL
EXIT
Press <TST TST> to view Test Functions.
NOTE: CO MEAS and CO REF will be artificially altered
to enforce a CO reading of 40.0 ppm.
All other Test Functions will report the correct
operational value.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...