Teledyne API – Model T300/T300M CO Analyzer
Calibration Procedures
229
9.7.2.3.
ACTIVATE CO2 SENSOR STABILITY FUNCTION
To change the stability test function from CO concentration to the CO
2
sensor output,
press:
Press EXIT 3
times to return
to SAMPLE
menu.
SETUP X.X
STABIL_GAS:CO2
CO CO2 O2
ENTR
EXIT
SAMPLE RANGE=50.0 PPM CO= XX.XX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG ALRM
EXIT
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
<PREV
NEXT>
JUMP
EDIT PRNT EXIT
SETUP X.X
2) STABIL_GAS=CO
<PREV NEXT> JUMP
EDIT
PRNT EXIT
SETUP X.X
STABIL_GAS:CO
CO
CO2
O2
ENTR EXIT
Continue pressing NEXT until ...
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
CO
2
and O
2
options only
appear if
associated
sensors are
installed.
Note
Use the same procedure to reset the STB test function to CO when the
CO2 calibration procedure is complete.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...