Setup Menu
Teledyne API – Model T300/T300M CO Analyzer
122
5.9.3.1.
ANALOG OUTPUT VOLTAGE / CURRENT RANGE SELECTION
In its standard configuration, each of the analog outputs is set to output a 0–5 VDC
signals. Several other output ranges are available. Each range has is usable from -5% to
+ 5% of the rated span.
Table 5-6: Analog Output Voltage Ranges
RANGE NAME
RANGE SPAN
MINIMUM OUTPUT
MAXIMUM OUTPUT
0.1V
0-100 mVDC
-5 mVDC
105 mVDC
1V
0-1 VDC
-0.05 VDC
1.05 VDC
5V
0-5 VDC
-0.25 VDC
5.25 VDC
10V
0-10 VDC
-0.5 VDC
10.5 VDC
•
The default offset for all VDC ranges is 0-5 VDC.
CURR
0-20 mA
0 mA
20 mA
•
While these are the physical limits of the current loop modules, typical applications use 2-20 or 4-20 mA for the lower and upper
limits. Please specify desired range when ordering this option.
•
The default offset for all current ranges is 0 mA.
•
Current outputs are available only on A1-A3.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...