Setup Menu
Teledyne API – Model T300/T300M CO Analyzer
108
5.4.5.
DILUTION RATIO (OPTION)
This feature is a optional software utility that allows the user to compensate for any
dilution of the sample gas that may occur before it enters the sample inlet. Typically this
occurs in continuous emission monitoring (CEM) applications where the sampling
method used to remove the gas from the stack dilutes it.
Using the dilution ratio option is a 4-step process:
1. Select the appropriate units of measure (see Section 5.4.4).
2. Select the reporting range mode and set the reporting range upper limit (see Section
•
Ensure that the upper span limit entered for the reporting range is the maximum
expected concentration of the
UNDILUTED
gas.
3. Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluent and 1
part of sample gas):
SAMPLE
RANGE=50.0 PPM CO=XX.XX
<TST
TST> CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS
CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
DIL
EXIT
SETUP X.X
O2 DIL FACTOR:1.0 Gain
0
0
0
0
1
.0
ENTR EXIT
Toggle these buttons to
select dilution ratio gain
factor for CO gas.
Default = 1 (e.g. 1:1).
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
SETUP X.X
CO DIL FACTOR:1.0 Gain
0
0
0
0
1
.0
ENTR EXIT
Toggle these buttons to select dilution
ratio gain factor for CO
2
or O
2
gas.
NOTE
Display only appears if the optional
CO
2
or O
2
Sensor is installed.
4. Calibrate the analyzer.
•
Make sure that the calibration span gas is either supplied through the same dilution
system as the sample gas or has an appropriately lower actual concentration.
EXAMPLE: If the reporting range limit is set for 100 ppm and the dilution ratio
of the sample gas is 20 gain, either:
•
a span gas with the concentration of 100 ppm can be used if the span gas passes
through the same dilution steps as the sample gas, or;
•
a 5 ppm span gas must be used if the span gas
IS NOT
routed through the dilution
system.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...