Teledyne API – Model T300/T300M CO Analyzer
Setup Menu
105
5.4.3.3.
AUTO RANGE MODE (AUTO)
In
AUTO
range mode, the analyzer automatically switches the reporting range between
two user-defined ranges (low and high).
•
The unit will switch from low range to high range when the CO
2
concentration
exceeds 98% of the low range span.
•
The unit will return from high range back to low range once both the CO
2
concentration falls below 75% of the low range span.
In
AUTO
Range Mode the instrument reports the same data in the same range on both
the
A1
and
A2
outputs and automatically switches both outputs between ranges as
described above. Also the
RANGE
test function displayed on the front panel will be
replaced by two separate functions:
•
RANGE1
: The
LOW
range setting for all analog outputs.
•
RANGE2
: The
HIGH
range setting for all analog outputs.
The high/low range status is also reported through the external, digital status bits (See
Section 3.3.1.4).
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...