Getting Started
Teledyne API - Model T200H/T200M Operation Manual
50
VACUUM
PRESS URE
SENSO R
SAMPL E
PRESS URE
SENSO R
O
3
FL
O
W
SEN
SO
R
FLOW PRESSURE
SENSOR PCA
INSTRUMENT CHASSIS
OZONE DRYER
NO/NO
X
VALVE
AUTOZERO
VALVE
O
3
Destruct
O
3
GENERATOR
EXHAUST
GAS
OUTLET
Fil
te
r
SAMPLE
GAS
INLET
ZERO GAS
INLET
SPAN GAS
INLET
SAMPLE/ CAL
VALVE
ZERO/SPAN
VALVE
Orifice Dia.
0.004"
PUMP
NO
X
E
xh
au
st
S
c
ru
b
b
e
r
EXH
AU
ST
MAN
IF
O
L
D
Orifice Dia.
0.007"
Orifice Dia.
0.007"
REACTION
CELL
PMT
NC
COM
NO
NC
COM
NO
NC
COM
NO
NC
COM
NO
NO
2
Converter
Zero/Span
Valve Option
Figure 3-20:
T200M – Internal Pneumatics with Ambient Zero-Span Valve Option 50A
Table 3-12: Zero/Span Valve States
MODE
VALVE
CONDITION
SAMPLE
Sample/Cal
Open to sample gas inlet
Zero/Span
Open to zero air inlet
ZERO
CALIBRATION
Sample/Cal
Open to zero/span inlet (activated)
Zero/Span
Open to zero air inlet
SPAN
CALIBRATION
Sample/Cal
Open to zero/span inlet (activated)
Zero/Span
Open to span gas inlet / IZS gas (activated)
The state of the zero/span valves can also be controlled:
•
Manually from the analyzer’s front panel by using the
SIGNAL I/O
controls located
under the
DIAG
Menu (Section 4.13.2),
•
By activating the instrument’s AutoCal feature (Section 5.8),
•
Remotely by using the external digital control inputs (Section 4.15.1.2) or Ethernet
option.
•
Remotely through the RS-232/485 serial I/O ports (Section 4.11).
All supply lines should be vented outside of the analyzer’s enclosure. In order to
prevent back-diffusion and pressure drop effects, these vent lines should be between 2
and 10 meters in length.