3.4.5. CURRENT LOOP ANALOG OUTPUTS (OPT 41) SETUP
This option adds isolated, voltage-to-current conversion circuitry to the analyzer’s
analog outputs. This option may be ordered separately for the first three of the analog
outputs and can be installed at the factory or added later. Call Teledyne API sales for
pricing and availability.
The current loop option can be configured for any output range between 0 and 20 mA
(for example 0-20, 2-20 or 4-20 mA). Information on calibrating or adjusting these
outputs can be found in Section 4.13.6.3.
Figure 3-9:
Current Loop Option Installed on the Motherboard
CAUTION – Avoid Warranty Invalidation
Printed Circuit Assemblies (PCAs) are sensitive to electro-static discharges too small to be
felt by the human nervous system. Failure to use Electro-Static Discharge (ESD) protection
when working with electronic assemblies will void the instrument warranty. Refer to the
manual, Fundamentals of ESD, PN 04786, which can be downloaded from our website at
under Help Center > Product Manuals in the Special Manuals
section.