APPENDIX A-3: Warnings and Test Measurements
Table A-2: Warning Messages
Name
1
Message Text
Description
Warnings
WSYSRES
SYSTEM RESET
Instrument was power-cycled or the CPU
was reset.
WDATAINIT
DATA INITIALIZED
Data storage was erased.
WCONFIGINIT
CONFIG INITIALIZED
Configuration storage was reset to factory
configuration or erased.
WNOXALARM1
9
NOX ALARM 1 WARN
NO
X
concentration alarm limit #1
exceeded
WNOXALARM2
9
NOX ALARM 2 WARN
NO
X
concentration alarm limit #2
exceeded
WNOALARM1
9
NO ALARM 1 WARN
NO concentration alarm limit #1 exceeded
WNOALARM2
9
NO ALARM 2 WARN
NO concentration alarm limit #2 exceeded
WNO2ALARM1
9
NO2 ALARM 1 WARN
NO
2
concentration alarm limit #1
exceeded
WNO2ALARM2
9
NO2 ALARM 2 WARN
NO
2
concentration alarm limit #2
exceeded
WO2ALARM1
5+9
O2 ALARM 1 WARN
O
2
concentration alarm limit #1 exceeded
WO2ALARM2
5+9
O2 ALARM 2 WARN
O
2
concentration alarm limit #2 exceeded
WCO2ALARM1
8+9
CO2 ALARM 1 WARN
CO
2
concentration alarm limit #1
exceeded
WCO2ALARM2
8+9
CO2 ALARM 2 WARN
CO
2
concentration alarm limit #2
exceeded
WSAMPFLOW
SAMPLE FLOW WARN
Sample flow outside of warning limits.
WOZONEFLOW
OZONE FLOW WARNING
Ozone flow outside of warning limits.
WOZONEGEN
OZONE GEN OFF
Ozone generator is off. This is the only
warning message that automatically
clears itself. It clears itself when the ozone
generator is turned on.
WRCELLPRESS
RCELL PRESS WARN
Reaction cell pressure outside of warning
limits.
WBOXTEMP
BOX TEMP WARNING
Chassis temperature outside of warning
limits.
WRCELLTEMP
RCELL TEMP WARNING
Reaction cell temperature outside of
warning limits.
WMANIFOLDTEMP
4
MANIFOLD TEMP WARN
Bypass or dilution manifold temperature
outside of warning limits.
WCO2CELLTEMP
8
CO2 CELL TEMP WARN
CO
2
sensor cell temperature outside of
warning limits.
WO2CELLTEMP
5
O2 CELL TEMP WARN
O
2
sensor cell temperature outside of
warning limits.
WIZSTEMP
IZS TEMP WARNING
IZS temperature outside of warning limits.
WCONVTEMP
CONV TEMP WARNING
Converter temperature outside of warning
limits.
WPMTTEMP
PMT TEMP WARNING
PMT temperature outside of warning
limits.
WAUTOZERO
WPREREACT
11
AZERO WRN XXX.X MV
PRACT WRN XXX.X MV
11
Auto-zero reading above limit. Value
shown in message indicates auto-zero