Teledyne API - Model T200H/T200M Operation Manual
Principles of Operation
297
Table 8-5: Typical Thermocouple Settings
TC
TYPE
TERMINATION
TYPE
OUTPUT
SCALE TYPE
JUMPER
BETWEEN
PINS
USED ON
JUMPER
COLOR
INPUT TC1 (J15)
K
GROUNDED
5mV / °C
2 – 12
4 – 14
T200H/M with Mini HiCon Converter
BROWN
K
ISOLATED
5mV / °C
2 – 12
4 – 14
5 – 15
T200H/M with Mini HiCon Converter
GREY
K
ISOLATED
10mV / °C
4 – 14
5 – 15
T200H/M models with Moly Converter
PURPLE
J
ISOLATED
10mV / °C
1 – 11
3 – 13
5 – 15
T200H/M models with Moly Converter
RED
J
GROUNDED
10mV / °C
1 – 11
3 – 13
T200H/M models with Moly Converter
GREEN
8.5.7.4. Valve Control
The relay board also hosts two valve driver chips, each of which can drive up four valves.
The main valve assembly in the T200H/M is the NO/NO
X
- Auto-zero solenoid valve
assembly mounted right in front of the NO
2
converter housing. These two valves are
actuated with 12 V supplied from the relay board and driven by the CPU through the I
2
C
bus.
A second set of valves may be installed if the zero/span valve is enabled in the analyzer.
Specialty manifold valves may be present in the analyzer.
07270D DCN7141