Teledyne API - Model T200H/T200M Operation Manual
Troubleshooting & Service
241
7.5.4.2. AC Configuration – Standard Heaters (JP2)
Power configuration for the AC the standard heaters is set using Jumper set JP2 (see
Figure 7-4 for the location of JP2).
Table 7-4: Power Configuration for Standard AC Heaters (JP2)
LINE VOLTAGE
JUMPER
COLOR
HEATER(S)
JUMPER
BETWEEN
PINS
FUNCTION
110 VAC / 115 VAC
50Hz & 60 Hz
WHITE
Reaction Cell / Sample
Chamber Heaters
1 to 8
Common
2 to 7
Neutral to Load
Mini Hi-Con
Converter
3 to 10
Common
4 to 9
Neutral to Load
Moly Converter
3 to 10
Common
4 to 9
Neutral to Load
Bypass Manifold
1
5 to 12
Common
6 to 11
Neutral to Load
220 VAC / 240 VAC
50Hz & 60 Hz
BLUE
Reaction Cell / Sample
Chamber Heaters
1 to 7
Load
Hi Concentration
Converter
3 to 9
Load
Moly Converter
3 to 9
Load
Bypass Manifold
1
5 to 11
Load
1
Bypass manifold is built into the reaction cell
Figure 7-13:
Typical Set Up of AC Heater Jumper Set (JP2)