Operating Instructions
Teledyne API - Model T200H/T200M Operation Manual
136
To Turn the over-range feature on or off, press:
From the
AIO CONFIGURATION SUBMENU
(in the SETUP DIAG menu)
DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL
<SET
SET>
EDIT
EXIT
DIAG AIO DATA_OUT_2 OVERRANGE: ON
<SET
SET>
EDIT
EXIT
DIAG AIO DATA_OUT_2 RANGE: 5V
SET>
EDIT
EXIT
DIAG AIO DATA_OUT_2 OVERRANGE: ON
ON
ENTR EXIT
DIAG AIO DATA_OUT_2 OVERRANGE: OFF
OFF
ENTR
EXIT
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
Toggle this button
to turn the Over-
Range feature ON
or OFF
Continue pressing
SET>
until you reach the
output to be configured
DIAG AIO AOUTS CALIBRATED: NO
SET>
CAL
EXIT
4.13.5.4. Adding a Recorder Offset to an Analog Output
Some analog signal recorders require that the zero signal is significantly different from
the baseline of the recorder in order to record slightly negative readings from noise
around the zero point. This can be achieved in the T200H/M by defining a zero offset, a
small voltage (e.g., 10% of span).