Getting Started
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
58
VENT
here if input
is pressurized
SAMPLE Gas
Source
PUMP
Chassis
SAMPLE
EXHAUST
SPAN1
Enclo
sure Wall
SPAN2/
VENT
Cal
ibrated N
O
X
at
HIGH
S
pan
Concen
tration
ZERO AIR
Zero Air Scrubber
Figure 3-24:
Gas Line Connection w/Ambient Zero/Pressurized Span Valves (OPT 50B)
S
AMPLE
G
AS
S
OURCE
Attach a sample inlet line to the SAMPLE inlet fitting.
Use PTFE tubing; minimum O.D ¼”.
Sample Gas pressure must equal ambient atmospheric pressure (29.92 in-Hg).
In applications where the sample gas is received from a pressurized manifold, a
vent must be placed on the sample gas line. This vent line must be:
no more than 10 meters long
vented outside the shelter or immediate area surrounding the instrument
C
ALIBRATION
G
AS
S
OURCES
SPAN GAS
Attach a gas line from the pressurized source of calibration gas (e.g. a
bottle of NISTSRM gas) to the SPAN1 inlet. Use PTFE tubing, minimum
O.D ¼”.
ZERO AIR
(the dual-stage zero Air Scrubber makes zero air)
V
ENTING
Attach a line to the SPAN2/VENT outlet. It should be:
¼” PTFE tubing
vented outside the enclosure
not less than 2 meters in length
not greater than 10 meters in length
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...