Instrument Maintenance
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
218
Figure 11-4:
Zero Air Scrubber Assembly
11.3.7.1. CHANGING THE EXTERNA
L SCRUBBER’S DFU FILTER
There is also a DFU filter on the inlet of the external zero air scrubber that is included in
several of the optional calibration valve packages.
To change this filter:
1. Disconnect the tube and fitting from one end and remove the filter from the scrubber
canister.
2. Insert a new filter and reattach the tubing.
3. Ensure that the small arrow embedded on the filter points in flow direction, i.e., to
analyzer.
06858E DCN7057