Instrument Maintenance
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
212
7. Check the O
3
flow rate; it should be around 80 cm³/min ± 15.
8. Check the RCEL pressure.
It should be the same value as before.
9. Refer to Section 11.3.12to leak check after installing the new DFU particle filter.
11.3.3. CHANGING THE OZONE CLEANSER CHEMICAL
The ozone (O
3
) cleanser is located next to the O
3
generator (see Figure 3-5) and cleans
the O
3
stream from solid and liquid contaminants that are created inside the O
3
generator. The content of the ozone cleanser needs periodical exchange according to
Table 11-1. A rebuild kit is available from the factory (see Appendix B of this manual
lists the part numbers).
To change the ozone cleanser chemical, follow these steps:
1. Turn of power to the analyzer and pump. Remove the analyzer cover and locate the
O
3
filter in the front of the analyzer next to the O
3
generator.
2. Use a 7/16” wrench to remove both pieces of 1/8” male nut with tubing from the
NPT fittings.
Figure 11-3:
Ozone Cleanser Assembly
06858E DCN7057