9.4. MANUAL CALIBRATION AND CAL CHECKS WITH THE
VALVE OPTIONS INSTALLED
There are a variety of valve options available on the T200 for handling calibration gases
(see Section 1.3 for descriptions of each).
Generally performing calibration checks and zero/span point calibrations on analyzers
with these options installed is similar to the methods discussed in the previous sections.
The primary differences are:
On instruments with Z/S valve options, zero air and span gas is supplied to the
analyzer through other gas inlets besides the sample gas inlet.
The zero and span calibration operations are initiated directly and independently
with dedicated buttons (
CALZ
&
CALS
).
9.4.1. SETUP FOR CALIBRATION USING VALVE OPTIONS
Each of the various calibration valve options requires a different pneumatic setup that is
dependent on the exact nature and number of valves present. Refer to the following
diagrams for information on each or these valve sets.